Reliability and accuracy analysis of a new semiautomatic radiographic measurement software in adult scoliosis.

Spine (Phila Pa 1976)

Department of Mechanical Engineering, École Polytechnique de Montréal, Downtown Station, Montréal, Québec, Canada.

Published: May 2011

AI Article Synopsis

  • The study evaluates a new radiographic software for measuring scoliosis in adults, focusing on its accuracy and reliability among different observers.
  • Previous methods relied heavily on repeated measures by experienced users, which can overlook the need for thorough accuracy assessments in scoliosis evaluations.
  • Results showed that most measurements had strong reliability, though there were issues with measuring sacral obliquity, highlighting the need for further improvement in that area.

Article Abstract

Study Design: Radiographic software measurement analysis in adult scoliosis.

Objective: To assess the accuracy as well as the intra- and interobserver reliability of measuring different indices on preoperative adult scoliosis radiographs using a novel measurement software that includes a calibration procedure and semiautomatic features to facilitate the measurement process.

Summary Of Background Data: Scoliosis requires a careful radiographic evaluation to assess the deformity. Manual and computer radiographic process measures have been studied extensively to determine the reliability and reproducibility in adolescent idiopathic scoliosis. Most studies rely on comparing given measurements, which are repeated by the same user or by an expert user. A given measure with a small intra- or interobserver error might be deemed as good repeatability, but all measurements might not be truly accurate because the ground-truth value is often unknown. Thorough accuracy assessment of radiographic measures is necessary to assess scoliotic deformities, compare these measures at different stages or to permit valid multicenter studies.

Methods: Thirty-four sets of adult scoliosis digital radiographs were measured two times by three independent observers using a novel radiographic measurement software that includes semiautomatic features to facilitate the measurement process. Twenty different measures taken from the Spinal Deformity Study Group radiographic measurement manual were performed on the coronal and sagittal images. Intra- and intermeasurer reliability for each measure was assessed. The accuracy of the measurement software was also assessed using a physical spine model in six different scoliotic configurations as a true reference.

Results: The majority of the measures demonstrated good to excellent intra- and intermeasurer reliability, except for sacral obliquity. The standard variation of all the measures was very small: ≤ 4.2° for Cobb angles, ≤ 4.2° for the kyphosis, ≤ 5.7° for the lordosis, ≤ 3.9° for the pelvic angles, and ≤5.3° for the sacral angles. The variability in the linear measurements (distances) was <4 mm. The variance of the measures was 1.7 and 2.6 times greater, respectively, for the angular and linear measures between the inter- and intrameasurer reliability. The image quality positively influenced the intermeasurer reliability especially for the proximal thoracic Cobb angle, T10-L2 lordosis, sacral slope and L5 seating. The accuracy study revealed that on average the difference in the angular measures was < 2° for the Cobb angles, and < 4° for the other angles, except T2-T12 kyphosis (5.3°). The linear measures were all <3.5 mm difference on average.

Conclusion: The majority of the measures, which were analyzed in this study demonstrated good to excellent reliability and accuracy. The novel semiautomatic measurement software can be recommended for use for clinical, research or multicenter study purposes.

Download full-text PDF

Source
http://dx.doi.org/10.1097/BRS.0b013e3181f0825aDOI Listing

Publication Analysis

Top Keywords

measurement software
16
radiographic measurement
12
adult scoliosis
12
measurement
8
intra- interobserver
8
software includes
8
semiautomatic features
8
features facilitate
8
facilitate measurement
8
intra- intermeasurer
8

Similar Publications

For the purpose of this study, four natural rock samples-namely, diorite, granodiorite, tonalite, and granite-are being investigated about their radiation attenuation. The elemental composition of the rocks was obtained through Energy dispersive X-ray spectroscopy (EDX) which examines the microstructural and localized area elemental analyses of the four rock samples. A Monte Carlo simulation (MCNP) was used to determine and evaluate the investigated samples.

View Article and Find Full Text PDF

Background: The aim of this study was to evaluate the correlation of the volume and minimum axial area (MAA) measurements between different upper and lower boundaries used for oropharyngeal airway assessment.

Methods: Cone Beam Computed Tomography (CBCT) scans of 49 subjects taken for pre-orthognathic surgical planning were obtained retrospectively from the archives (n = 49; 32 females, 17 males; mean age = 20.9 ± 5.

View Article and Find Full Text PDF

Purpose: This study aimed to compare different treatment modalities to correct ill-fitted maxillary complete denture either by the conventional relining method or by scanning the relining impression and digitally construct a new denture regarding patient satisfaction, denture retention, and adaptation.

Materials And Methods: Twelve edentulous patients suffering from loose maxillary complete dentures were selected, dentures' borders and fitting surfaces were prepared, and relining impressions were taken, the impressions were scanned and the STL files were used for CAD/CAM milling ( computer aided designing/ computer aided manufacturing) of new maxillary dentures (Group A), then the relining impression went through the conventional laboratory steps to fabricate (Group B) maxillary dentures. Both groups were evaluated regarding patient satisfaction by a specially designed questionnaire, retention values were measured by a digital force gauge at denture insertion appointment and two weeks later, geomagic software was used to evaluate dentures adaptation to oral tissues.

View Article and Find Full Text PDF

Exploring policy processes against microbial threats in Iran: a qualitative policy analysis.

BMJ Open

January 2025

Department of Environmental Health Engineering, School of Public Health, Tehran University of Medical Sciences, Tehran, Iran (the Islamic Republic of).

Objectives: Microbial threats pose a growing concern worldwide. This paper reports the analysis of Iran's policy process against microbial threats.

Design: This is a qualitative study.

View Article and Find Full Text PDF

Background: For many colorectal cancer patients, primary surgery is the standard care of treatment. Further insights in perioperative care are crucial. The aim of this study is to assess the prognostic value of body composition for postoperative complications after laparoscopic and open colorectal surgery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!