Purpose: We evaluate the magnitude of metallic artifacts caused by various implantable cardiac pacing devices (without leads) on both attenuation maps (μ-maps) and positron emission tomography (PET) images using experimental phantom studies. We also assess the efficacy of a metal artifact reduction (MAR) algorithm along with the severity of artifacts in the presence of misalignment between μ-maps and PET images.
Methods: Four pacing devices including two pacemakers (pacemakers 1 and 2) and two cardiac resynchronization therapy (CRT) devices of pacemaker (CRT-P) and defibrillator (CRT-D) type were placed in three phantoms including a cylindrical Ge-68 phantom, a water-bath phantom and an anthropomorphic heart/thorax phantom. The μ-maps were derived from computed tomography (CT) images reconstructed using the standard method supplied by the manufacturer and those reconstructed using the MAR algorithm. In addition, the standard reconstructed CT images of the last two phantoms were manually misaligned by 10 mm along the patient's axis to simulate misalignment between CT and PET images.
Results: The least and severest artifacts produced on both μ-maps and PET images of the Ge-68 phantom were induced by CRT-P and pacemaker 1 devices, respectively. In the water-bath phantom, CRT-P induced 17.5% over- and 9.2% underestimation of tracer uptake whereas pacemaker 1 induced 69.6% over- and 65.7% underestimation. In the heart/thorax phantom representing a pacemaker-bearing patient, pacemaker 1 induced 41.8% increase and 36.6% decrease in tracer uptake and attenuation coefficients on average in regions corresponding to bright and dark streak artifacts, respectively. Statistical analysis revealed that the MAR algorithm was successful in reducing bright streak artifacts, yet unsuccessful for dark ones. In the heart/thorax phantom, the MAR algorithm reduced the overestimations to 4.4% and the underestimations to 35.5% on average. Misalignment between μ-maps and PET images increased the peak of pseudo-uptake by approximately 20%.
Conclusions: This study demonstrated that, depending on their elemental composition, different implantable cardiac pacing devices result in varying magnitudes of metal artifacts and thus pseudo-uptake on PET images. The MAR algorithm was not successful in compensating for underestimations which calls for a more efficient algorithm. The results showed that misalignments between PET and CT images render metal-related pseudo-uptake more severe.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s11307-010-0467-x | DOI Listing |
Front Pharmacol
December 2024
Department of Nuclear Medicine, The First Affiliated Hospital, Jinan University, Guangzhou, China.
Objective: To optimize the automated radiosynthesis of the purinergic ion channel receptor 7 (P2X7R) imaging agent F-JNJ64413739 and evaluate its potential for brain imaging in osteoporotic model rats.
Methods: A more electron-deficient nitropyridine was employed as the labeling precursor to facilitate the F-labeling. The radiosynthesis was conducted on an AllinOne synthesis module, and followed by purification via high-performance liquid chromatography (HPLC).
Front Oncol
December 2024
Department of Ultrasound, Hebei Medical University Third Hospital, Shijiazhuang, Hebei, China.
Desmoplastic malignant peritoneal mesothelioma (DMPM) is an extremely rare and aggressive subtype of sarcomatoid malignant mesothelioma, originating from the mesothelial lining of body cavities. It is characterized by significant local invasiveness and poor prognosis. The nonspecific symptoms of DMPM often result in delayed diagnosis.
View Article and Find Full Text PDFNucl Med Rev Cent East Eur
December 2024
Department of Radiology & Nuclear Medicine, Sultan Qaboos Comprehensive Cancer Care, and Research Center, Muscat, Oman.
Background: In radioembolization therapy for hepatic malignancies, the accurate estimation of lung shunt fraction (LSF) is crucial to minimize the risk of radiation-induced pneumonitis and fibrosis due to hepatopulmonary shunting of yttrium-90 (90Y)-microspheres. This study aimed to compare the accuracy and precision of LSF estimation using technetium-99m macroaggregated albumin single photon emission computed tomography ([99mTc]Tc-MAA SPECT) LSF, [99mTc]Tc-MAA planar LSF, and 90Y PET LSF in patients undergoing 90Y-radioembolization.
Material And Methods: A retrospective study was conducted involving 15 patients diagnosed with hepatocellular carcinoma (HCC) or liver metastases and planned to undergo transarterial radioembolization with 90Y SirSpheres after multidisplinary team discussion.
Eur J Nucl Med Mol Imaging
December 2024
Department of Nuclear Medicine, Peking University First Hospital, No. 8 Xishiku Str., Xicheng Dist, Beijing, 100034, China.
Purpose: CD38 is a glycoprotein highly specific to multiple myeloma (MM). Therapeutics using antibodies targeting CD38 have shown promising efficacy. However, the efficient stratification of patients who may benefit from daratumumab (Dara) therapy and timely monitoring of therapeutic responses remain significant clinical challenges.
View Article and Find Full Text PDFEur J Nucl Med Mol Imaging
December 2024
Division of Urology, Department of Surgery, University of Alberta, Edmonton, Canada.
Purpose: Fluorine-18 prostate-specific membrane antigen-1007 positron emission tomography/computed tomography (F-PSMA-1007 PET/CT) has been shown to be superior to multiparametric magnetic resonance imaging (MRI) for the locoregional staging of intermediate-risk and high-risk prostate tumors. This study aims to evaluate whether it is also superior in estimating tumor parameters, such as three-dimensional spatial localization and volume.
Methods: 134 participants underwent F-PSMA-1007 PET/CT and MRI prior to radical prostatectomy as part of the validating paired-cohort Next Generation Trial (NCT05141760).
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!