Purpose: To compare the course of intraocular pressure (IOP) during corneal flap preparation using four different femtosecond lasers in porcine globes.
Methods: Forty-eight (12 in each group) enucleated globes were successfully cannulated through the optic nerve. Intraocular pressure was measured continuously through the cannula during a normal lamellar flap creation (regular procedure) using four femtosecond lasers (IntraLase, Abbott Medical Optics; VisuMax, Carl Zeiss Meditec AG; Femtec, Technolas Perfect Vision; and Femto LDV, Ziemer Ophthalmic Systems AG). In an additional measurement (worst-case procedure), the patient interface was pressed against the globe with increasing force until the applanation maneuver was automatically aborted by those devices capable of doing so.
Results: During the regular procedure, the maximum IOP reached was 135±16 mmHg when using the Intra-Lase, 65±20 mmHg with the VisuMax, 205±32 mmHg with the Femtec, and 184±28 mmHg with the Femto LDV. During the worst-case procedure, a maximum IOP of 260±53 mmHg was reached with the IntraLase, 105±13 mmHg with the VisuMax, and 248±51 mmHg with the Femtec.
Conclusions: There is considerable variation in IOP among the tested femtosecond lasers during a regular lamellar flap creation and during the worst-case procedure. The VisuMax femtosecond laser seems to cause the lowest IOP rise in both settings.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3928/1081597X-20101210-01 | DOI Listing |
Materials (Basel)
December 2024
College of Mechanical Engineering, Zhejiang University of Technology, Hangzhou 310023, China.
As an environment-friendly material, graphene oxide nanosheet can effectively improve the polishing surface quality of single crystal diamond workpieces. However, the lubricating and chemical effects of graphene oxide nanosheets have an uncertain impact on the polishing material removal rate. In this paper, the graphene oxide-enhanced hybrid slurry was prepared with good stability.
View Article and Find Full Text PDFCells
December 2024
Institute of Solid State Physics, Bulgarian Academy of Sciences, Tzarigradsko Chaussee 72, 1784 Sofia, Bulgaria.
Laser-induced photothermal therapy using gold nanoparticles (AuNPs) has emerged as a promising approach to cancer therapy. However, optimizing various laser parameters is critical for enhancing the photothermal conversion efficacy of plasmonic nanomaterials. In this regard, the present study investigates the photothermal effects of dodecanethiol-stabilized hydrophobic ultrasmall spherical AuNPs (TEM size 2.
View Article and Find Full Text PDFBioengineering (Basel)
November 2024
Singapore National Eye Center, Singapore 168751, Singapore.
We describe retrospectively the indications and outcomes of nine patients who present with varying degrees of deep posterior stromal scarring or endothelial failure following deep anterior lamellar keratoplasty (DALK). These patients underwent a surgical strategy coined Intraoperative Optical Coherence Tomography Guided Femtosecond Laser-Assisted Descemet Membrane Endothelial Keratoplasty (iFAD). This strategy can be used to address suboptimal visual outcomes following primary DALK.
View Article and Find Full Text PDFHeliyon
January 2025
Laser and Plasma Research Institute, Shahid Beheshti University, Tehran, Iran.
The MKN45 cell line, a type of gastric cancer cell, exhibits resistance to chemotherapy agents through various mechanisms. Curcumin and noscapine, two plant-derived anticancer compounds, exhibit selective cytotoxicity towards cancer cells. However, their bioavailability is poor both in vitro and in vivo.
View Article and Find Full Text PDFLasers Surg Med
January 2025
ViaLase Inc., Aliso Viejo, California, USA.
Objectives: Femtosecond laser image guided high precision trabeculotomy (FLigHT) is a novel open-angle glaucoma treatment. The procedure non-invasively creates aqueous humor (AH) drainage channel from the anterior chamber (AC) to Schlemm's canal (SC) through the trabecular meshwork (TM) to decrease intraocular pressure (IOP). The purpose of this study was to develop a 3D finite element model (FEM) of the FLigHT procedure and to simulate clinical results for different drainage channel cross-sectional areas.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!