Introduction: Volar fixed-angle plating is a popular treatment for unstable distal radius fractures. Despite the availability of plating systems for treating distal radius fractures, little is known about the mechanical properties of multidirectional fixed-angle plates. The aim of this study was to compare the primary fixation stability of three possible screw configurations in a distal extra-articular fracture model using a multidirectional fixed-angle plate with metaphyseal cancellous screws distally.

Material And Methods: Eighteen Sawbones radii (Sawbones, Sweden, model# 1027) were used to simulate an extra-articular distal radius fracture according to AO/OTA 23 A3. Plates were fixed to the shaft with one non-locking screw in the oval hole and two locking screws as recommended by the manufacturer. Three groups (n = 6) were defined by screw configuration in the distal metaphyseal fragment: Group 1: distal row of screws only; Group 2: 2 rows of screws, parallel insertion; Group 3: 2 rows of screws, proximal screws inserted with 30° of inclination. Specimens underwent mechanical testing under axial compression within the elastic range and load controlled between 20 N and 200 N at a rate of 40 N/s. Axial stiffness and type of construct failure were recorded.

Results: There was no difference regarding axial stiffness between the three groups. In every specimen, failure of the Sawbone-implant-construct occurred as plastic bending of the volar titanium plate when the dorsal wedge was closed.

Discussion: Considering the limitations of the study, the recommendation to use two rows of screws or to place screws in the proximal metaphyseal row with inclination cannot be supported by our mechanical data.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00508-010-1488-9DOI Listing

Publication Analysis

Top Keywords

distal radius
16
rows screws
12
screws
9
volar fixed-angle
8
fixed-angle plating
8
locking screws
8
three screw
8
screw configurations
8
extra-articular fracture
8
fracture model
8

Similar Publications

Objective: Investigating the clinical efficacy of treating dorsally displaced distal radial double-column Die-punch fractures using a dorsal approach external fixator combined with Kirschner wires.

Methods: Retrospectively analyzed the clinical data of 15 patients with distal radial double-column Die-punch fractures treated with an external fixator combined with Kirschner wire between July 2020 and January 2023. There were 10 males and 5 females;6 cases on the left side and 9 on the right;age ranged from 22 to 76 years old.

View Article and Find Full Text PDF

Objective: Meta-analysis of the clinical efficacy of plate and external fixator fixation in the treatment of AO-C type distal radius fractures.

Methods: PubMed, Embase, Cochrane Medical Library, Web of Science, CNKI, Wanfang, VIP and SinoMed databases were searched for all literature on randomized controlled clinical trials of AO-C distal radius fractures. The search time limits were from each database.

View Article and Find Full Text PDF

Objective: To explore the accuracy of human-computer interaction software in identifying and locating type C1 distal radius fractures.

Methods: Based on relevant inclusion and exclusion criteria, 14 cases of type C1 distal radius fractures between September 2023 and March 2024 were retrospectively analyzed, comprising 3 males and 11 females(aged from 27 to 82 years). The data were assigned randomized identifiers.

View Article and Find Full Text PDF

Objective: To investigate the short-term clinical effect of closed reduction single arm external fixator combined with percutaneous needle fixation in the treatment of C1 distal radius fracture in elderly patients.

Methods: Between December 2022 and December 2023, a total of 60 elderly patients diagnosed with type C1 distal radius fractures were treated, comprising 9 males and 51 females. The age ranged from 65 to 84 years old, with an average of (72.

View Article and Find Full Text PDF

Objective: To compare the short-term clinical efficacy and safety of closed reduction with Kirschner wire fixation versus open reduction with plate fixation for treating osteoporotic Colles' fractures in middle-aged and elderly patients.

Methods: Between January 2018 and January 2023, 119 patients with Colles fractures were retrospectively analyzed, including 39 males and 80 females, aged from 48 to 74 years old with an average of(60.58±6.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!