Purpose: Second-generation FloTrac software has been shown to reliably measure cardiac output (CO) in cardiac surgical patients. However, concerns have been raised regarding its accuracy in vasoplegic states. The aim of the present multicenter study was to investigate the accuracy of the third-generation software in patients with sepsis, particularly when total systemic vascular resistance (TSVR) is low.

Methods: Fifty-eight septic patients were included in this prospective observational study in four university-affiliated ICUs. Reference CO was measured by bolus pulmonary thermodilution (iCO) using 3-5 cold saline boluses. Simultaneously, CO was computed from the arterial pressure curve recorded on a computer using the second-generation (CO(G2)) and third-generation (CO(G3)) FloTrac software. CO was also measured by semi-continuous pulmonary thermodilution (CCO).

Results: A total of 401 simultaneous measurements of iCO, CO(G2), CO(G3), and CCO were recorded. The mean (95%CI) biases between CO(G2) and iCO, CO(G3) and iCO, and CCO and iCO were -10 (-15 to -5)% [-0.8 (-1.1 to -0.4) L/min], 0 (-4 to 4)% [0 (-0.3 to 0.3) L/min], and 9 (6-13)% [0.7 (0.5-1.0) L/min], respectively. The percentage errors were 29 (20-37)% for CO(G2), 30 (24-37)% for CO(G3), and 28 (22-34)% for CCO. The difference between iCO and CO(G2) was significantly correlated with TSVR (r(2) = 0.37, p < 0.0001). A very weak (r(2) = 0.05) relationship was also observed for the difference between iCO and CO(G3).

Conclusions: In patients with sepsis, the third-generation FloTrac software is more accurate, as precise, and less influenced by TSVR than the second-generation software.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3028067PMC
http://dx.doi.org/10.1007/s00134-010-2098-8DOI Listing

Publication Analysis

Top Keywords

flotrac software
12
cardiac output
8
third-generation software
8
septic patients
8
patients sepsis
8
pulmonary thermodilution
8
ico cog2
8
difference ico
8
ico
7
software
6

Similar Publications

Small changes in the transducer position cause a systematic change in cardiac output readings: implications for clinical practice.

J Clin Monit Comput

October 2024

Department of Cardiothoracic- and Vascular Surgery, Aarhus University Hospital, Anesthesia Section, Palle Juul-Jensens Boulevard 99, Aarhus N, 8200, Denmark.

To systematically evaluate the effect of small changes in transducer position on key hemodynamic variables including CO generated by 4th generation FloTrac software. After cardiac surgery, cardiac output, mean arterial pressure, systemic vascular resistance, and stroke volume variation were measured with 4 generation Flotrac software. The transducer position was randomly placed at the midaxillary plane, 4 cm higher than the midaxillary plane or 4 cm lower than the midaxillary plane.

View Article and Find Full Text PDF

The pulmonary artery catheter (PAC)-despite its invasiveness-remains the gold standard for cardiac output (CO) monitoring. The FloTrac system, a less invasive hemodynamic monitor has been developed, which estimates CO using arterial pressure waveform analysis without external calibration. Recently, an upgraded version of FloTrac system with improved algorithm to follow changes in vascular resistance was introduced into the market.

View Article and Find Full Text PDF

The pulmonary artery catheter (PAC) is considered the gold standard for cardiac index monitoring. Recently new and less invasive methods to assess cardiac performance have been developed. The aim of our study was to assess the reliability of a non-invasive monitor utilizing bioreactance (Starling SV) and a non-calibrated mini-invasive pulse contour device (FloTrac/EV1000, fourth-generation software) compared to bolus thermodilution technique with PAC (TDCO) during off-pump coronary artery bypass surgery (OPCAB).

View Article and Find Full Text PDF

Objective: To determine the accuracy of a continuous cardiac output monitor (FloTrac sensor) for measuring cardiac index in children with congenital heart disease undergoing cardiac catheterization. Cardiac index is a critical hemodynamic parameter measured during catheterizations in children with congenital heart disease. This has been challenging to measure accurately and many clinicians rely on predictive equations for calculating cardiac index.

View Article and Find Full Text PDF

Background: Passive leg raising (PLR) is a so called self-volume challenge used to test for fluid responsiveness. Changes in cardiac output (CO) or stroke volume (SV) measured during PLR are used to predict the need for subsequent fluid loading. This requires a device that can measure CO changes rapidly.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!