A study of peak skin doses (PSDs) during neuroradiology and cardiology interventional procedures has been carried out using Gafchromic XR-RV2 film. Use of mosaics made from squares held in cling film has allowed doses to the head to be mapped successfully. The displayed cumulative air kerma (CAK) has been calibrated in terms of cumulative entrance surface dose (CESD) and results indicate that this can provide a reliable indicator of the PSD in neuroradiology. Results linking PSD to CESD for interventional cardiology were variable, but CAK is still considered to provide the best option for use as an indicator of potential radiation-induced effects. A CESD exceeding 3 Gy is considered a suitable action level for triggering follow-up of patients in neuroradiology and cardiology for possible skin effects. Application of dose action levels defined in this way would affect 8% of neurological embolisation procedures and 5% of cardiology ablation and multiple stent procedures at the hospitals where the investigations were carried out. A close relationship was observed between CESD and dose-area product (DAP) for particular types of procedure, and DAPs of 200-300 Gy cm(2) could be used as trigger levels where CAK readings were not available. The DAP value would depend on the mean field size and would need to be determined for each application.