Micro and nano-structure hydroxyapatite samples were synthesized via several different methods. The samples were characterised utilising the Fourier transmission infra-red, scanning electron microscope and X-ray diffraction methods, to find out the structure most similar to human tooth enamel, and the best method was found. The electron paramagnetic resonance (EPR) signals of the gamma-irradiated samples were measured using an EPR spectrometer system. A calibration curve was established by irradiation of the samples at four doses of 50-500 mGy. The parameters of the calibration curve, slope and intercept with dose axis are determined by linear regression analysis. This calibration curve can be used for human tooth enamel for retrospective dosimetry purposes.