Background: Glycated hemoglobin (A1C) monitoring is an integral component of diabetes management. This study was conducted to evaluate the performance of the A1CNow® SELFCHECK device when used by lay users and health care professionals (HCPs) to measure A1C.
Methods: Subjects performed two A1CNow SELFCHECK finger-stick self-tests followed by a finger-stick test of the subject's blood by a HCP. The primary endpoint assessed accuracy of the subject and HCP A1CNow SELFCHECK readings. Secondary endpoints included precision, comprehension of instructional material (written material±DVD), and product satisfaction. For accuracy comparison, a venous blood sample was drawn from each subject and tested by laboratory (TOSOH) analysis. Subject comprehension of product instructional material was evaluated via first-time failure (FTF) rate as recorded by the HCP, and subject satisfaction was assessed through written survey.
Results: A total of 110 subjects with (n=93) and without (n=17) diabetes participated. Of 177 subject A1C values, 165 (93.2%) were within the acceptable range of ±13.5% of the laboratory reference value and considered accurate. Regression analysis showed good correlation of subject values to laboratory and HCP results (R2=0.93 for both). The average within-subject coefficient of variation was 4.57% (n=74). The FTF rates with and without instructional DVD were 11.3% (n=56) and 39.6% (n=54), respectively. Subjects with diabetes/prediabetes overwhelmingly indicated that they were "very" to "extremely" likely (93.5%) to discuss their home A1C results with their HCP.
Conclusions: Lay users found the A1CNow SELFCHECK easy to use, and both lay users and HCPs were able to measure A1C accurately.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3005061 | PMC |
http://dx.doi.org/10.1177/193229681000400625 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!