As new functionality is added to the implantable devices, their power requirements also increase. Such power requirements make it hard for keeping such implants operational for long periods by non-rechargeable batteries. This result in a need for frequent surgeries to replace these batteries. Rechargeable batteries can satisfy the long-term power requirements of these new functions. To minimize the discomfort to the patients, the recharging of the batteries should be as infrequent as possible. Traditional battery charging methods have low battery charging efficiency. This means they may limit the amount of charge that can be delivered to the device, speeding up the depletion of the battery and forcing frequent recharging. In this paper, we evaluate the suitability of a state-of-the-art general purpose charging method called current-pumped battery charger (CPBC) for implant applications. Using off-the-shelf components and with minimum optimization, we prototyped a proof-of-concept transcutaenous battery charger based on CPBC and show that the CPBC can charge a 100 mAh battery transcutaneously within 137 minutes with at most 2.1°C increase in tissue temperature even with a misalignment of 1.3 cm in between the coils, while keeping the battery charging efficiency at 85%.