Within-subjects comparison of the HiRes and Fidelity120 speech processing strategies: speech perception and its relation to place-pitch sensitivity.

Ear Hear

Department of Communication Sciences and Disorders, College of Behavioral and Community Sciences, University of South Florida, Tampa, FL 33620-8100, USA.

Published: June 2011

Objectives: Previous studies have confirmed that current steering can increase the number of discriminable pitches available to many cochlear implant (CI) users; however, the ability to perceive additional pitches has not been linked to improved speech perception. The primary goals of this study were to determine (1) whether adult CI users can achieve higher levels of spectral cue transmission with a speech processing strategy that implements current steering (Fidelity120) than with a predecessor strategy (HiRes) and, if so, (2) whether the magnitude of improvement can be predicted from individual differences in place-pitch sensitivity. A secondary goal was to determine whether Fidelity120 supports higher levels of speech recognition in noise than HiRes.

Design: A within-subjects repeated measures design evaluated speech perception performance with Fidelity120 relative to HiRes in 10 adult CI users. Subjects used the novel strategy (either HiRes or Fidelity120) for 8 wks during the main study; a subset of five subjects used Fidelity120 for three additional months after the main study. Speech perception was assessed for the spectral cues related to vowel F1 frequency, vowel F2 frequency, and consonant place of articulation; overall transmitted information for vowels and consonants; and sentence recognition in noise. Place-pitch sensitivity was measured for electrode pairs in the apical, middle, and basal regions of the implanted array using a psychophysical pitch-ranking task.

Results: With one exception, there was no effect of strategy (HiRes versus Fidelity120) on the speech measures tested, either during the main study (N = 10) or after extended use of Fidelity120 (N = 5). The exception was a small but significant advantage for HiRes over Fidelity120 for consonant perception during the main study. Examination of individual subjects' data revealed that 3 of 10 subjects demonstrated improved perception of one or more spectral cues with Fidelity120 relative to HiRes after 8 wks or longer experience with Fidelity120. Another three subjects exhibited initial decrements in spectral cue perception with Fidelity120 at the 8-wk time point; however, evidence from one subject suggested that such decrements may resolve with additional experience. Place-pitch thresholds were inversely related to improvements in vowel F2 frequency perception with Fidelity120 relative to HiRes. However, no relationship was observed between place-pitch thresholds and the other spectral measures (vowel F1 frequency or consonant place of articulation).

Conclusions: Findings suggest that Fidelity120 supports small improvements in the perception of spectral speech cues in some Advanced Bionics CI users; however, many users show no clear benefit. Benefits are more likely to occur for vowel spectral cues (related to F1 and F2 frequency) than for consonant spectral cues (related to place of articulation). There was an inconsistent relationship between place-pitch sensitivity and improvements in spectral cue perception with Fidelity120 relative to HiRes. This may partly reflect the small number of sites at which place-pitch thresholds were measured. Contrary to some previous reports, there was no clear evidence that Fidelity120 supports improved sentence recognition in noise.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3042485PMC
http://dx.doi.org/10.1097/AUD.0b013e3181fb8390DOI Listing

Publication Analysis

Top Keywords

fidelity120
16
speech perception
16
place-pitch sensitivity
16
fidelity120 relative
16
relative hires
16
main study
16
spectral cues
16
vowel frequency
16
hires fidelity120
12
spectral cue
12

Similar Publications

Previous work showed that the Fidelity120 processing strategy provides better spectral sensitivity, while the HiResolution processing strategy can deliver more detailed temporal information for Advanced Bionics cochlear implant users. The goal of this study was to develop a new sound processing strategy by maximizing the spectral benefit of Fidelity120 and the temporal benefit of HiResolution to improve both aspects of hearing. Using acoustic simulations of Fidelity120 and HiResolution strategies, a dual-processing strategy was created by combining Fidelity120 in the low frequency channels and HiResolution in the high frequency channels.

View Article and Find Full Text PDF

Within-subjects comparison of the HiRes and Fidelity120 speech processing strategies: speech perception and its relation to place-pitch sensitivity.

Ear Hear

June 2011

Department of Communication Sciences and Disorders, College of Behavioral and Community Sciences, University of South Florida, Tampa, FL 33620-8100, USA.

Objectives: Previous studies have confirmed that current steering can increase the number of discriminable pitches available to many cochlear implant (CI) users; however, the ability to perceive additional pitches has not been linked to improved speech perception. The primary goals of this study were to determine (1) whether adult CI users can achieve higher levels of spectral cue transmission with a speech processing strategy that implements current steering (Fidelity120) than with a predecessor strategy (HiRes) and, if so, (2) whether the magnitude of improvement can be predicted from individual differences in place-pitch sensitivity. A secondary goal was to determine whether Fidelity120 supports higher levels of speech recognition in noise than HiRes.

View Article and Find Full Text PDF

Sensitivity of psychophysical measures to signal processor modifications in cochlear implant users.

Hear Res

April 2010

V.M. Bloedel Hearing Research Center, Department of Otolaryngology, University of Washington, Seattle, WA 98195, United States.

Experienced users of the Clarion cochlear implant were tested acutely with the HiResolution (HiRes) and HiRes Fidelity120 (F120) processing strategies. Three psychophysically-based tests were used including spectral-ripple discrimination, Schroeder-phase discrimination and temporal modulation detection. Three clinical outcome measures were used including consonant-nucleus-consonant (CNC) word recognition in quiet, word recognition in noise and the clinical assessment of music perception (CAMP).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!