Purpose: To evaluate a model of mechanically measuring resistance to tearing of a continuous curvilinear capsulorhexis (CCC) using the entire capsular bag of fresh human cadaver eyes isolated from the eyes after complete evacuation.
Setting: John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.
Design: Experimental study.
Methods: After the cornea and iris were removed, a 5.0 to 5.5 mm anterior CCC was created. The nucleus was hydroexpressed and the capsular bag evacuated by irrigation/aspiration. A pair of metal shoetree-shaped fixtures, designed based on human lens geometric dimensions, were implanted separately in the capsular bag and assembled together with a screw nut. After complete zonulectomy, the fixture-capsular bag assembly was removed from the eye and loaded onto a mechanical tester. The fixtures were separated at a velocity of 7.0 mm/min in 0.15 μm intervals to stretch the CCC to its rupture point. Rupture load (N) and extension were measured and graphed.
Results: Testing of 23 donor eyes a mean of 69.04 hours ± 22.72 (SD) after death showed the following mean values: CCC diameter, 5.3 ± 0.12 mm; load, 0.39 ± 0.16 N; extension at CCC tearing, 5.85 ± 1.17 mm. There was a moderately strong negative correlation between donor age and load (P = .0018).
Conclusions: Previous mechanized methods of assessing CCC strength used excised anterior capsules or partially phacoemulsified crystalline lenses, yielding lower tension strength values. This force-displacement method may facilitate assessment of small differences between anterior capsulotomy techniques.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcrs.2010.05.026 | DOI Listing |
J Fr Ophtalmol
January 2025
Service d'ophtalmologie, centre hospitalier de Boulogne-sur-Mer, rue Jacques-Monod, 59000 Boulogne-sur-Mer, France. Electronic address:
Biomedicines
January 2025
Department of Ophthalmology, UHC Sestre Milosrdnice, 10000 Zagreb, Croatia.
We describe a novel experimental device, the VaMa (Vatavuk and Marić) artificial intraocular lens (IOL) capsule, and a method that enables all IOL types to be implanted in the bag. We present the application of the device and the procedure in patients with aphakia and native capsule damage and without efficient capsular support. The VaMa device and the method facilitate IOL exchange due to refractive errors and, in the case of their invention, the implementation of superior IOLs in the future.
View Article and Find Full Text PDFAm J Ophthalmol
January 2025
Raghudeep Eye Hospital, Laden Cataract and IOL research centre, Ahmedabad, India 380052.
Purpose: To evaluate demographic profile, risk factors and outcomes following intraocular lens (IOL) exchange surgery for late decentration of IOLs in the dead bag syndrome DESIGN: Prospective, interventional case series PARTICIPANTS: 46 eyes (43 patients) with late decentration/dislocation of IOLs and spontaneous posterior capsule rupture in a clear and relatively intact capsular bag. None of the eyes had significant zonular weakness.
Methods: Study conducted at Raghudeep Eye Hospital, India.
GMS Ophthalmol Cases
December 2024
Department of Ophthalmology, Disha Eye Hospital, Siliguri, India.
Background: Pseudophakic cystoid macular edema (CME) following primary anterior-chamber intraocular lens (ACIOL) implantations is commonly seen. Intravitreal triamcinolone acetonide (IVTA) injections have shown significant improvement in visual acuity and retinal thickness in refractory pseudophakic CME. Pseudohypopyon following IVTA injection is a known entity.
View Article and Find Full Text PDFSci Rep
December 2024
Visual Optics Lab Antwerp (VOLANTIS), Faculty of Medicine and Health Sciences, University of Antwerp, Antwerp, Belgium.
In cataract surgery, post-surgical stability of the intraocular lens plays a major role. This study aims to explore how the size and decentration of the capsulorhexis affect intraocular lens decentration and tilt by using numerical methods. Finite element models included zonules, ciliary body, capsular bag, and an IOL with two open-loop haptics were built.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!