Background: We studied long-term right ventricular (RV) pacing threshold (RVPT) behavior in patients consecutively implanted with pacemakers capable of automatic output reprogramming tracked by automatic RV threshold measurement (automatic verification of capture [AVC]).
Methods: All the patients had state-of-the art steroid-eluting bipolar pacing leads and were RV-paced by an AVC algorithm from the three American manufacturers. Follow-up occurred twice in the first year after implantation, then yearly until approaching elective replacement indicator.
Results: Three hundred and twenty-one patients aged 73 ± 12 years were observed for 49 ± 26 months on average. At implantation, RVPT was 0.54 ± 0.2 V at 0.4 ms at an average 774 ± 217 Ω impedance. Forty-one of the 321 patients (12.8%) had a permanent RVPT increase above 1.5 V at 0.4 ms: RVPT was between 1.6 and 2.5 V in 29 of 321 (9%) patients, whereas it was between 2.6 and 3.5 V in seven of 321 (2.2%) patients, and >3.5 V in five of 321 (1.5%) patients. No exit block occurred because of automatic RV output adjustment by AVC algorithms. No predictor of RVPT increase was found at multivariable analysis. The maximum RVPT increase occurred within 12 months from implantation in 19 of 321 (5.9%) patients, between the first and the second year in 12 of 321 (3.7%), between the second and the sixth year in eight of 321 (2.5%), and after the sixth year in two of 321 (0.6%).
Conclusion: Despite technologic improvement in lead manufacturing, long-term increase of the RVPT occurs in about 13% of patients, possibly representing a serious safety issue in 3.7% when 2.5 V at 0.4 ms is exceeded. AVC algorithms can improve patients' safety by automatic tailoring of the pacing output to threshold fluctuations, while maximizing device longevity.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1540-8159.2010.02915.x | DOI Listing |
J Interv Card Electrophysiol
September 2019
Section of Electrophysiology, Division of Cardiology, Beijing Friendship Hospital Affiliated to Capital Medical University, No. 95 Yongan Road, Xi-cheng District, Beijing, 100050, China.
Purpose: Right ventricular pacing threshold (RVPT) may rise over time accompanied by the increased use of implantable cardiac pacemakers. However, risk factors for permanent RVPT increase are not fully clarified in patients without definite lead fracture and dislodgment. We aimed to evaluate the long-term performance of RV pacing leads and identify risk factors associated with the occurrence of permanent RVPT increase in this population.
View Article and Find Full Text PDFHeart Vessels
September 2017
Department of Anesthesiology and Intensive Care Medicine, Osaka University, 2-2 Yamadaoka, Suita, Osaka, Japan.
Rapid ventricular pacing (RVP) is used during transcatheter aortic valve implantation (TAVI). RVP disturbs myocardial oxygen balance, and when prolonged, it may cause procedure-related myocardial injury (PMI). This study investigated whether a longer duration of RVP increased the occurrence of PMI or worsened long-term mortality after TAVI.
View Article and Find Full Text PDFPacing Clin Electrophysiol
January 2011
S.Orsola-Malpighi Hospital, Institute of Cardiology, University of Bologna, Bologna, Italy.
Background: We studied long-term right ventricular (RV) pacing threshold (RVPT) behavior in patients consecutively implanted with pacemakers capable of automatic output reprogramming tracked by automatic RV threshold measurement (automatic verification of capture [AVC]).
Methods: All the patients had state-of-the art steroid-eluting bipolar pacing leads and were RV-paced by an AVC algorithm from the three American manufacturers. Follow-up occurred twice in the first year after implantation, then yearly until approaching elective replacement indicator.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!