Tip geometry and placement of rotary blood pump inflow and outflow cannulae influence the dynamics of flow within the ventricle and aortic branch. Cannulation, therefore, directly influences the potential for thrombus formation and end-organ perfusion during ventricular assist device (VAD) support or cardiopulmonary bypass (CPB). The purpose of this study was to investigate the effect of various inflow/outflow cannula tip geometries and positions on ventricular and greater vessel flow patterns to evaluate ventricular washout and impact on cerebral perfusion. Transparent models of a dilated cardiomyopathic ventricle and an aortic branch were reconstructed from magnetic resonance imaging data to allow flow measurements using particle image velocimetry (PIV). The contractile function of the failing ventricle was reproduced pneumatically, and supported with a rotary pump. Flow patterns were visualized around VAD inflow cannulae, with various tip geometries placed in three positions in the ventricle. The outflow cannula was placed in the subclavian artery and at several positions in the aorta. Flow patterns were measured using PIV and used to validate an aortic flow computational fluid dynamic study. The PIV technique indicated that locating the inflow tip in the left ventricular outflow tract improved complete ventricular washout while the tip geometry had a smaller influence. However, side holes in the inflow cannula improved washout in all cases. The PIV results confirmed that the positioning and orientation of the outflow cannula in the aortic branch had a high impact on the flow pattern in the vessels, with a negative blood flow in the right carotid artery observed in some cases. Cannula placement within the ventricle had a high influence on chamber washout. The positioning of the outflow cannula directly influences the flow through the greater vessels, and may be responsible for the occasional reduction in cerebral perfusion seen in clinical CPB.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1525-1594.2010.01098.x | DOI Listing |
JACC Case Rep
December 2024
Jesselson Integrated Heart Center, The Eisenberg R&D Authority, Shaare Zedek Medical Center, Faculty of Medicine, The Hebrew University of Jerusalem, Jerusalem, Israel.
Objective: This study sought to present the endovascular approach of transcatheter aspiration using the FlowTriever (Inari Medical) aspiration system for high surgical risk patients with right-sided infective endocarditis.
Key Steps: General anesthesia and transesophageal echocardiogram guidance; ultrasonography-guided femoral vein access, preclosure sutures, and insertion of a 24-F sheath; insertion of straight 24-F aspiration cannula over a stiff wire, parked in the superior vena cava; introduction of a 20-F curved cannula inside the 24-F cannula to create a telescopic assembly; accurate positioning using the right ventricle inflow/outflow projection in biplane mode; adjustment of the curved cannula radius by sliding the inner cannula in and out inside the mother cannula; manual aspiration of the vegetation; Postaspiration transesophageal echocardiogram assessment.
Potential Pitfalls: Avoid leaflet and annular injury and account for potential embolization.
Plast Reconstr Surg Glob Open
January 2025
From the Division of Plastic and Reconstructive Surgery, Mayo Clinic, Rochester, MN.
Pyogenic flexor tenosynovitis (PFT), also known as septic or suppurative flexor tenosynovitis, is a closed-space infection of the hand's flexor tendon sheath that necessitates timely diagnosis and treatment. The treatment consists of antibiotic therapy often combined with prompt surgical treatment. The most common surgical approach is the closed irrigation technique, which involves inserting a 16-gauge angiocatheter in the proximal aspect of the flexor tendon sheath, leaving the distal end of the Brunner incision open during the irrigation process.
View Article and Find Full Text PDFAnn Surg Oncol
December 2024
Hepato-Pancreato-Biliary Surgery and Liver Transplantation, Pôle des Pathologies Hépatiques et Digestives, Hôpital de Hautepierre-Hôpitaux Universitaires de Strasbourg, Université de Strasbourg, Strasbourg, France.
Background: Total vascular exclusion (TVE) with liver hypothermic perfusion under venovenous bypass (VVB) is usually needed to perform hepatectomy with Inferior vena cava and hepatic veins resection-reconstruction. An alternative technique is represented by liver resection under intermittent pedicular clamping, IVC total clamping and VVB, without cold perfusion and liver outflow drainage through the VVB. PATIENTS AND METHODS: The patient is a 60-year-old woman with past medical history of right hepatectomy for leiomyosarcoma 14 years previously.
View Article and Find Full Text PDFIndian J Ophthalmol
December 2024
Department of Oculoplasty, Disha Eye Hospitals Pvt Ltd, Kolkata, West Bengal, India.
JTCVS Open
October 2024
Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio.
Objective: A less-invasive left atrial assist device (LAADx) is a novel and implantable, extracardiac blood pump concept, intended for the treatment of diastolic heart failure, represented by heart failure with preserved ejection fraction.
Methods: A mixed-flow pump was used as the working LAADx model. Its performance was evaluated at 3 speeds, using an in vitro pulsatile mock circulatory loop, with a pneumatic pump that can simulate diastolic heart failure conditions by adjusting the diastolic drive pressure.