Background: Test phantoms with simulated micro-calcifications of true calcium hydroxyapatite (CaHA) density were not available to validate advanced calcium scoring methods or plaque density measurements.
Objectives: We evaluated a coronary calcium scoring (CCS) test phantom containing very small CaHA microspheres and validated a new scoring method for measurements of plaque densities.
Methods: The semianthropomorphic CCS phantom was constructed with CaHA microspheres (volumes, 0.05-3.1 mm(3)) with the approximate density of biologic calcifications. QRM and CCS phantoms were scored with a new calibrated and automated calcium scoring method (N-vivo; Image Analysis). The densities of the microspheres and 609 individual patient plaques were measured.
Results: The range of measured densities of the CaHA microspheres was approximately equivalent to that measured in the patient coronary calcifications. The smallest microspheres scored with the calibrated/automated and the Agatston methods had volumes of 0.075 mm(3) and 0.27 mm(3), respectively. The standard deviations of the mass scores of the microspheres ranged from 0.02 to 0.17 mg with regression slope of 0.962 and R(2) = 0.997. The relationship of measured density to measured mass of the patient plaques was similar to that of the microspheres, suggesting that vascular calcifications are CaHA density.
Conclusions: The CaHA microspheres of the CCS test phantom were found to be representative in density and size of coronary calcifications. The measurements show that CT calcium scoring underestimates plaque density and greatly overestimates volume. The heterogeneity of calcium concentration densities measured in the patient plaques was due largely to CT scanner measurement errors.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcct.2010.08.004 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!