[Application of partial robust M-regression in noninvasive measurement of human blood glucose concentration with near-infrared spectroscopy].

Guang Pu Xue Yu Guang Pu Fen Xi

Key Laboratory of Precision Opto-mechatronics Technology, Ministry of Education, School of Instrument Science and Opto-Electronics Engineering, Beihang University, Beijing 100191, China.

Published: August 2010

In the study of non-invasive measurement of human blood glucose concentration with near-infrared spectroscopy, the partial robust M-regression (PRM) is proposed in the present paper to solve the robustness of calibration model affected by outliers existing in the spectra data set. While keeping the good properties of M-estimators if an appropriate weighting scheme is chosen, PRM inherits the speed of computation and easy realization of the iterative reweighted partial least squares (IRPLS) algorithm, but is robust to all types of outliers. With the pretreatment of spectra based on PRM, the root mean square error of prediction (RMSEP) of calibration model was presented and compared with partial least squares (PLS). Experimental results show that the robust calibration model PRM produces better prediction of glucose than the model of PLS when the components of the samples increase which is significant for non-invasive prediction of blood glucose levels.

Download full-text PDF

Source

Publication Analysis

Top Keywords

blood glucose
12
calibration model
12
partial robust
8
robust m-regression
8
measurement human
8
human blood
8
glucose concentration
8
concentration near-infrared
8
partial squares
8
[application partial
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!