Dosimetric effects of an air cavity for the SAVI partial breast irradiation applicator.

Med Phys

Department of Radiation Oncology, Washington University School of Medicine, St. Louis, Missouri 63110, USA.

Published: August 2010

Purpose: To investigate the dosimetric effect of the air inside the SAVI partial breast irradiation device.

Methods: The authors have investigated how the air inside the SAVI partial breast irradiation device changes the delivered dose from the homogeneously calculated dose. Measurements were made with the device filled with air and water to allow comparison to a homogenous dose calculation done by the treatment planning system. Measurements were made with an ion chamber, TLDs, and film. Monte Carlo (MC) simulations of the experiment were done using the EGSnrc suite. The MC model was validated by comparing the water-filled calculations to those from a commercial treatment planning system.

Results: The magnitude of the dosimetric effect depends on the size of the cavity, the arrangement of sources, and the relative dwell times. For a simple case using only the central catheter of the largest device, MC results indicate that the dose at the prescription point 1 cm away from the air-water boundary is about 9% higher than the homogeneous calculation. Independent measurements in a water phantom with a similar air cavity gave comparable results. MC simulation of a realistic multidwell position plan showed discrepancies of about 5% on average at the prescription point for the largest device.

Conclusions: The dosimetric effect of the air cavity is in the range of 3%-9%. Unless a heterogeneous dose calculation algorithm is used, users should be aware of the possibility of small treatment planning dose errors for this device and make modifications to the treatment delivery, if necessary.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.3457328DOI Listing

Publication Analysis

Top Keywords

air cavity
12
savi partial
12
partial breast
12
breast irradiation
12
treatment planning
12
dosimetric air
8
air inside
8
inside savi
8
dose calculation
8
prescription point
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!