Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Aims: Control of atrioventricular (AV) node conduction by means of high-frequency stimulation (HFS) of efferent AV node vagal stimulation (AVNS) fibres enables the ventricular rate (VR) to be modulated during atrial fibrillation (AF). The aims of this study were to verify, on 18-month follow-up, the reproducibility of the dromotropic effect obtained on implantation and the long-term reliability of the system in patients who received an implantable cardioverter-defibrillator (ICD) with a standard atrial lead positioned at a location suitable for AVNS.
Methods And Results: We enrolled 12 patients with paroxysmal or persistent AF who were candidates for ICD. The right atrium was mapped to locate the pacing site, and a transvenous screw-in lead was implanted in that region. The voltages required for VR modulation (25% VR reduction) and complete AV block at different pulse durations (from 0.1 to 0.5 ms) were recorded. Eleven out of 12 patients underwent 18-month follow-up examination. Atrial pacing parameters were adequate and did not differ from the baseline values (all P > 0.05): pacing threshold 0.9 ± 0.5 V (0.5 ms pulse duration) and impedance 556 ± 121 Ω, with P-wave amplitude of 1.6 ± 0.7 mV. High-frequency stimulation induced VR modulation in nine patients and complete AV block in eight patients at pulse durations ≥0.3 ms. No differences were observed in the voltages for VR modulation and complete AV block between implantation and 18-month examination (all P > 0.100).
Conclusion: Ventricular rate control during AF was obtained under HFS 18 months after implantation in patients with the atrial lead positioned at a location suitable for AVNS. The pacing outputs needed to achieve the dromotropic effect were comparable to those measured on implantation.