Background: The incidence, risk factors, and management of very high defibrillation thresholds (DFTs) during present-day implantable cardioverter defibrillator (ICD) testing are not well known.

Objectives: The purpose of this study was to assess (1) the incidence of very high DFTs and (2) the efficacy/safety of routinely adding a subcutaneous (SQ) array for these patients.

Methods: The study evaluated patients undergoing first-time ICD implantation at Southlake Regional Healthcare Centre from January 2006 to December 2007. All implanted ICDs had a maximal output of 35 J. Patients with DFTs greater than a 10-J safety margin from maximum output were considered to have very high readings and underwent SQ array insertion after other attempts at lowering DFT (group I). These patients were compared with the rest of the patients who had acceptable DFTs (group II) using both univariate and multivariate logistic regression analysis. Outcomes of array insertion were also assessed.

Results: A total of 313 patients underwent first-time ICD implantation during the analysis period. Of those, 16 (5.1%) had very high DFTs (group I). By univariate analysis, advanced New York Heart Association class (3 or 4), congestive heart failure hospitalization, non-ischemic cardiomyopathy, amiodarone use, implant of a biventricular device, and highest quartile of left ventricular (LV) chamber enlargement were all significant predictors of very high DFTs (p < 0.05). By multivariate analysis, only amiodarone use [odds ratio (OR) = 10.3, 95% confidence interval (95% CI) = 3.7-32.6] and being in the highest quartile for LV diastolic diameter [OR = 5.4, 95% CI = 1.4-20.8] predicted very high DFT. In all 16 cases, other methods to lower DFT prior to array insertion were attempted but failed for all patients: reversing shock polarity (n = 15), removing the superior vena cava coil (n = 14), reprogramming shock waveform (n = 9), and repositioning right ventricular lead (n = 9). Addition of the array successfully decreased DFT to within safety margin for all patients (33 ± 2 vs 21 ± 5 J, p = 0.02). Complication due to array insertion occurred in one patient (pneumothorax).

Conclusions: Very high DFTs occur in about 5% of patients undergoing ICD implantation and may be predicted by LV dilation and amiodarone use. SQ array insertion reliably corrects this problem over other interventions with a low rate of procedural complication.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10840-010-9511-7DOI Listing

Publication Analysis

Top Keywords

array insertion
12
icd implantation
12
high dfts
12
incidence high
8
high defibrillation
8
defibrillation thresholds
8
subcutaneous array
8
implantable cardioverter
8
cardioverter defibrillator
8
defibrillator icd
8

Similar Publications

In this paper, a multilayer monopulse antenna at Ku-Band with high efficiency, high power handling capability, high gain, 45° linear polarization and low sidelobe is presented. A new slot antenna is proposed as a radiating element based on a cavity-backed slot-coupled patch antenna. Using an enclosed cavity structure reduces coupling between antenna elements, thus increasing the antenna efficiency.

View Article and Find Full Text PDF

The nucleolus is a major subnuclear compartment where ribosomal DNA (rDNA) is transcribed and ribosomes are assembled. In addition, recent studies have shown that the nucleolus is a dynamic organizer of chromatin architecture that modulates developmental gene expression. rDNA gene units are assembled into arrays located in the p-arms of five human acrocentric chromosomes.

View Article and Find Full Text PDF

Improving Real-Time Feedback During Cochlear Implantation: The Auditory Nerve Neurophonic/Cochlear Microphonic Ratio.

Ear Hear

January 2025

Department of Otorhinolaryngology and Head and Neck Surgery, Radboud University Medical Centre, Donders Institute for Brain, Cognition and Behaviour, Nijmegen, The Netherlands.

Objectives: Real-time monitoring of cochlear function to predict the loss of residual hearing after cochlear implantation is now possible. Current approaches monitor the cochlear microphonic (CM) during implantation from the electrode at the tip of the implant. A drop in CM response of >30% is associated with poorer hearing outcomes.

View Article and Find Full Text PDF

Cochlear Implant Electrode Placement and Music Perception.

JAMA Otolaryngol Head Neck Surg

January 2025

Department of Hearing and Speech Sciences, Vanderbilt University Medical Center, Nashville, Tennessee.

Importance: Cochlear implants enable improvements in speech perception, but music perception outcomes remain variable. Image-guided cochlear implant programming has emerged as a potential programming strategy for increasing the quality of spectral information delivered through the cochlear implant to improve outcomes.

Objectives: To perform 2 experiments, the first of which modeled the variance in music perception scores as a function of electrode positioning factors, and the second of which evaluated image-guided cochlear implant programming as a strategy to improve music perception with a cochlear implant.

View Article and Find Full Text PDF

The mitochondrial outer membrane (OMM) β-barrel proteins link the mitochondrion with the cytosol, endoplasmic reticulum, and other cellular membranes, establishing cellular homeostasis. Their active insertion and assembly in the outer mitochondrial membrane is achieved in an energy-independent yet highly effective manner by the Sorting and Assembly Machinery (SAM) of the OMM. The core SAM constituent is the 16-stranded transmembrane β-barrel Sam50.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!