Purpose: To perform a preliminary evaluation of the image quality of pediatric technetium 99m ((99m)Tc) methylene diphosphonate (MDP) bone single photon emission computed tomography (SPECT) by using iterative reconstruction-ordered subset expectation maximization with three-dimensional resolution recovery (OSEM-3D)-and to assess whether any improvements with use of this technique could lead to a reduction in patient dose or a shortening in imaging time.

Materials And Methods: Institutional advisory board approval was obtained for this investigation. Fifty (99m)Tc-MDP SPECT studies of the spine were evaluated (36 female and 14 male patients; mean age, 15.5 years). Each study was acquired by using a dual-detector camera, with each detector rotating 360°. By using filtered back projection (FBP) and OSEM-3D, images were reconstructed from data generated by both detectors. Likewise, OSEM-3D was used to reconstruct data from a single detector simulating half the administered radiopharmaceutical activity. Two nuclear medicine physicians, blinded to the patient data, reviewed the images for image quality in four different categories by using a four-point scale: artifacts (category 1), lesions (category 2), noise (category 3), and image sharpness (category 4).

Results: Compared with FBP, images reconstructed by using OSEM-3D with one or two detectors showed significant improvement in image quality with regard to lesion detection, noise level, and image sharpness (P < .02, .01, and .001, respectively). With OSEM-3D, no significant differences were observed when either one or two detectors were used.

Conclusion: Improved image quality of skeletal SPECT with either a 50% reduction in radiation dose or a 50% reduction in acquisition time or combination of the two can be achieved by using OSEM-3D.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiol.10100102DOI Listing

Publication Analysis

Top Keywords

image quality
16
subset expectation
8
expectation maximization
8
resolution recovery
8
images reconstructed
8
image sharpness
8
50% reduction
8
image
6
osem-3d
5
pediatric 99mtc-mdp
4

Similar Publications

Abnormal network homogeneity in patients with bipolar disorder in attention network.

Brain Imaging Behav

January 2025

Key Laboratory of Adolescent Cyberpsychology and Behavior (Ministry of Education), Wuhan, China.

Bipolar disorder (BD) is a complex psychiatric condition marked by significant mood fluctuations that deeply affect quality of life. Understanding the neural mechanisms underlying BD is critical for improving diagnostic accuracy and developing more effective treatments. This study utilized resting-state functional magnetic resonance imaging (rs-fMRI) to investigate functional connectivity within the ventral and dorsal attention networks in 52 patients with BD and 51 healthy controls.

View Article and Find Full Text PDF

Deep Learning Superresolution for Simultaneous Multislice Parallel Imaging-Accelerated Knee MRI Using Arthroscopy Validation.

Radiology

January 2025

From the Department of Radiology, Division of Musculoskeletal Radiology, NYU Grossman School of Medicine, 660 1st Ave, 3rd Fl, Rm 313, New York, NY 10016 (S.S.W., J.V., R.K., E.H.P., J.F.); Department for Diagnostic and Interventional Radiology, Eberhard Karls University Tübingen, University Hospital Tübingen, Tübingen, Germany (S.S.W.); Department of Radiology, University Hospital Basel, Basel, Switzerland (J.V.); Department of Radiology, Hospital do Coraçao, São Paulo, Brazil (T.C.R.); Academic Surgical Unit, South West London Elective Orthopaedic Centre (SWLEOC), London, United Kingdom (D.D.); Department of Radiology, Balgrist University Hospital, Zurich, Switzerland (B.F.); Department of Radiology, Jeonbuk National University Hospital, Jeonju, Republic of Korea (E.H.P.); Research Institute of Clinical Medicine of Jeonbuk National University, Biomedical Research Institute of Jeonbuk National University Hospital, Jeonju, Republic of Korea (E.H.P.); Medscanlagos Radiology, Cabo Frio, Brazil (A.S.); Centre for Data Analytics, Bond University, Gold Coast, Australia (S.E.S.); Siemens Healthineers AG, Erlangen, Germany (I.B.); and Siemens Medical Solutions USA, Malvern, Pa (G.K.).

Background Deep learning (DL) methods can improve accelerated MRI but require validation against an independent reference standard to ensure robustness and accuracy. Purpose To validate the diagnostic performance of twofold-simultaneous-multislice (SMSx2) twofold-parallel-imaging (PIx2)-accelerated DL superresolution MRI in the knee against conventional SMSx2-PIx2-accelerated MRI using arthroscopy as the reference standard. Materials and Methods Adults with painful knee conditions were prospectively enrolled from December 2021 to October 2022.

View Article and Find Full Text PDF

Compressed ultrafast photography (CUP) is a high-speed imaging technique with a frame rate of up to ten trillion frames per second (fps) and a sequence depth of hundreds of frames. This technique is a powerful tool for investigating ultrafast processes. However, since the reconstruction process is an ill-posed problem, the image reconstruction will be more difficult with the increase of the number of reconstruction frames and the number of pixels of each reconstruction frame.

View Article and Find Full Text PDF

A symmetrical dual-D and dual-core single-mode fiber surface plasmon resonance (SPR) liquid sensor is designed for biological detection. The dual-core design optimizes the transmission path, improves the momentum matching between free electrons and photons, and facilitates bidirectional coupling, consequently amplifying the SPR effect and enabling sensitive monitoring of the refractive index changes of biological solutions. In this structure, a gold wire is placed in the middle of the polished surface of the double-D-shaped single-mode fiber (SMF) to produce high-quality free electrons and promote the mode-coupling excitation of the SPR effect.

View Article and Find Full Text PDF

Objectives: We evaluated the performance of a novel flow cell morphology analyzer AUTION EYE AI-4510 for counting particles in urine.

Methods: Analytical performance was assessed according to the EFLM European Urinalysis Guideline 2023. Trueness was compared by analyzing 1.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!