Background: Point-of-care D-dimer tests have recently been introduced to enable rapid exclusion of deep venous thrombosis (DVT) without the need to refer a patient for conventional laboratory-based D-dimer testing. Before implementation in practice, however, the diagnostic accuracy of each test should be validated.
Methods: We analyzed data of 577 prospectively identified consecutive primary care patients suspected to have DVT, who underwent 5 point-of-care D-dimer tests-4 quantitative (Vidas®, Pathfast™, Cardiac®, and Triage®) and 1 qualitative (Clearview Simplify®)-and ultrasonography as the reference method. We evaluated the tests for the accuracy of their measurements and submitted a questionnaire to 20 users to assess the user-friendliness of each test.
Results: All D-dimer tests showed negative predictive values higher than 98%. Sensitivity was high for all point-of-care tests, with a range of 0.91 (Clearview Simplify) to 0.99 (Vidas). Specificity varied between 0.39 (Pathfast) and 0.64 (Clearview Simplify). The quantitative point-of-care tests showed similar and high discriminative power for DVT, according to calculated areas under the ROC curves (range 0.88-0.89). The quantitative Vidas and Pathfast devices showed limited user-friendliness for primary care, owing to a laborious calibration process and long analyzer warm-up time compared to the Cardiac and Triage. For the qualitative Clearview Simplify assay, no analyzer or calibration was needed, but interpretation of a test result was sometimes difficult because of poor color contrast.
Conclusions: Point-of-care D-dimer assays show good and similar diagnostic accuracy. The quantitative Cardiac and Triage and the qualitative Clearview Simplify D-dimer seem most user-friendly for excluding DVT in the doctor's office.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1373/clinchem.2010.147892 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!