Comparison of the Sensititre YeastOne® dilution method with the Clinical Laboratory Standards Institute (CLSI) M27-A3 microbroth dilution reference method for determining MIC of eight antifungal agents on 102 yeast strains.

Pathol Biol (Paris)

UMR 145 (IRD/UM1), UFR sciences pharmaceutiques et biologiques, 15, avenue Charles-Flahault, BP 14491, 34093 Montpellier cedex 5, France.

Published: February 2011

The Clinical Laboratory Standards Institute ([CLSI] formerly NCCLS) reference broth microdilution testing method (protocol M27-A3) was compared with a commercially available methods (Sensititre YeastOne(®)) by testing two quality control strains and 102 isolates of Candida sp. and Cryptococcus sp. against fluconazole, itraconazole, ketoconazole, posaconazole, voriconazole, flucytosin, amphotericin B and caspofungin. Minimal inhibitory concentrations (MIC) endpoints were determined after 24h of incubation for Sensititre YeastOne(®) and after 24 and 48 h for CLSI microdilution method. Essential agreements between methods vary from 70.6 to 92.2%. Categorical agreements vary from 94.1% for 5FC to 72.6% for AMB. Sensititre YeastOne(®) reading appears to be useful for avoiding very major errors and this confirms the interest of this method for evaluating new antifungals activity in vitro.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.patbio.2010.07.020DOI Listing

Publication Analysis

Top Keywords

sensititre yeastone®
16
clinical laboratory
8
laboratory standards
8
standards institute
8
method
5
comparison sensititre
4
yeastone®
4
yeastone® dilution
4
dilution method
4
method clinical
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!