In vivo precision of the GE Lunar iDXA densitometer for the measurement of total body composition and fat distribution in adults.

Eur J Clin Nutr

DXA Unit, Carnegie Faculty of Sport and Education, Carnegie Research Institute, Leeds Metropolitan University, Headingley Campus, Leeds, UK.

Published: January 2011

In vivo precision for body composition measurements using dual energy X-ray absorptiometry (DXA; GE Lunar iDXA, GE Healthcare, Bucks, UK) was evaluated in 52 men and women, aged 34.8 (s.d. 8.4; range 20.1-50.5) years, body mass index (25.8 kg/m(2); range 16.7-42.7 kg/m(2)). Two consecutive total body scans (with re-positioning) were conducted. Precision was excellent for all measurements, particularly for total body bone mineral content and lean tissue mass (root mean square 0.015 and 0.244 kg; coefficients of variation (CV) 0.6 and 0.5%, respectively). Precision error was CV 0.82% for total fat mass and 0.86% for percentage fat. Precision was better for gynoid (root mean square 0.397 kg; CV 0.96%) than for android fat distribution (root mean square 0.780 kg, CV 2.32%). There was good agreement between consecutive measurements for all measurements (slope (s.e.) 0.993-1.002; all R(2) = 0.99). The Lunar iDXA provided excellent precision for total body composition measurements. Research into the effect of body size on the precision of DXA body fat distribution measurements is required.

Download full-text PDF

Source
http://dx.doi.org/10.1038/ejcn.2010.190DOI Listing

Publication Analysis

Top Keywords

total body
16
lunar idxa
12
body composition
12
fat distribution
12
root square
12
vivo precision
8
body
8
composition measurements
8
precision
6
measurements
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!