Validation of a 3-dimensional laser body scanner for assessment of waist and hip circumference.

J Am Coll Nutr

Department of Nutritional Sciences, The University of Texas at Austin, 1 University Station, A2703, Austin, TX 78712, USA.

Published: June 2010

Objectives: To evaluate the reliability and validity of a 3-dimensional laser body scanner for estimation of waist and hip circumferences and waist:hip ratio.

Methods: Seventy women were evaluated for waist and hip circumference and waist:hip ratio via laser scanner and tape measure. In a subset of 34 participants, 8 repeated measures of laser scanning were performed for reproducibility analysis. Validity of the instrument was assessed by regression and Bland-Altman comparison of measures of waist and hip circumferences and waist:hip ratio to tape measure.

Results: Reproducibility analysis showed little difference between within-subjects measurements of circumferences (intraclass correlation coefficient ≥0.992, p < 0.01). Evaluation of waist and hip circumferences measured by body scanning did not differ significantly from tape measure (p > 0.05). Bland-Altman analysis showed no bias between laser scanning and tape measure.

Conclusion: These findings indicate that the 3-dimensional laser body scanner is a reliable and valid technique for the estimation of waist and hip circumferences as compared with tape measure. This instrument is promising as a quick and simple method of body circumference analysis.

Download full-text PDF

Source
http://dx.doi.org/10.1080/07315724.2010.10719832DOI Listing

Publication Analysis

Top Keywords

waist hip
24
hip circumferences
16
3-dimensional laser
12
laser body
12
body scanner
12
tape measure
12
hip circumference
8
estimation waist
8
circumferences waisthip
8
waisthip ratio
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!