Background: RTVue spectral-domain optical coherence tomography (OCT) is a new, ultra high-speed and high-resolution instrument, potentially to measure the presence of glaucoma or its progression accurately. The objective of this study was to evaluate its reproducibility of retinal nerve fiber layer (RNFL) thickness and optic nerve head (ONH) measurements in normal and glaucoma eyes.
Methods: This study was an observational clinical study. One eye was selected randomly from each of 89 normal individuals and 63 glaucoma patients in a range of severity. RNFL thickness and ONH were measured 3 times on the same day to determine intrasession variability. The same instrument was used by the same operator for all scans. Intrasession within-subject standard deviation (Sw), precision (1.96 x Sw), coefficient of variation (CVw, 100 x Sw/overall mean), and intraclass coefficient (ICC) were calculated to evaluate reproducibility.
Results: RTVue OCT demonstrated double hump patterns in the RNFL profiles. High reproducibility was observed in all ONH parameters. For normal eyes, the value of ICC ranged between 0.98 and 1.00. For eyes with different extent of glaucoma, it ranged between 0.94 and 1.00. High reproducibility was also observed in RNFL thickness measurements. The values of ICC for averaged RNFL thickness ranged between 0.95 and 1.00 in all cases. For regional parameters, it ranged from 0.94 to 0.98 for normal eyes, 0.94 to 1.00 for mild glaucoma eyes, 0.87 to 1.00 for moderate glaucoma eyes, and 0.77 to 0.97 for severe glaucoma eyes. The nasal regions of severe glaucoma appeared to be most variable, as nasal lower region and inferior nasal region had the ICC values of 0.77 and 0.87.
Conclusion: Reproducibility of RTVue RNFL and ONH measurements was excellent in normal and glaucoma groups.
Download full-text PDF |
Source |
---|
Clin Ophthalmol
January 2025
Department of Ophthalmology and Visual Sciences, W. K. Kellogg Eye Center, University of Michigan Medical School, Ann Arbor, MI, USA.
Purpose: To understand patient experiences using the iCare HOME tonometer through assessing ease of use, device usage patterns, proficiency, and patient comments.
Methods: We conducted a prospective survey-based cohort study of 19 patients (35 eyes) with glaucoma or glaucoma suspect diagnoses. Patients received training on home tonometer usage and were loaned the device for one week to measure intraocular pressure at home multiple times daily.
Clin Ophthalmol
January 2025
Alcon, Inc, Fort Worth, TX, USA.
Purpose: To determine the effect on intraocular pressure (IOP) of switching to a once-daily netarsudil/latanoprost fixed dose combination (FDC) from various topical treatment regimens including latanoprost monotherapy or latanoprost combined with other IOP-lowering agents for the treatment of open-angle glaucoma or ocular hypertension.
Methods: A total of 136 participants enrolled. Eligible participants were aged ≥18 years and had a current diagnosis of open-angle glaucoma or ocular hypertension.
Clin Ophthalmol
January 2025
Department of Ophthalmology, Manhattan Eye, Ear, and Throat Hospital/Northwell Health, New York, NY, USA.
Purpose: To assess the diagnostic capability of pattern electroretinography (PERG) and varying circumpapillary optical coherence tomography (OCT) scan diameters in glaucoma suspects (GS).
Methods: This is a prospective, cross-sectional study. Circumpapillary retinal nerve fiber layer thickness (RNFLT) was measured using spectral domain OCT in 49 eyes from 26 patients (36 normal, 13 GS) in three circle diameters (3.
Clin Ophthalmol
January 2025
Department of Ophthalmology, Saitama Medical Center, Saitama Medical University, Kawagoe-shi, Saitama, Japan.
Purpose: Among patients with angle-closure glaucoma, it is common to have a short-axial eye, which also makes it difficult to select an appropriate intraocular lens. Previous studies have focused on the ocular biometry of the long-axial eye, whereas only a few reports have focused on the short-axial eye. This study aimed to clarify the characteristics of the short-axial eye on ocular biometry among the elderly Japanese.
View Article and Find Full Text PDFCureus
December 2024
Cornea and Refractive Surgery, Al-Shifa Trust Eye Hospital, Rawalpindi, PAK.
Background: Glaucoma, particularly open-angle glaucoma (OAG), is a leading cause of irreversible blindness, associated with optic nerve damage, retinal ganglion cell death, and visual field defects. Corneal biomechanical properties and cellular components, such as corneal nerve and keratocyte densities assessed by in vivo confocal microscopy (IVCM), may serve as biomarkers for glaucoma progression. This study aimed to explore the relationship between corneal nerve parameters, keratocyte density, and optical coherence tomography (OCT)-derived retinal nerve fiber layer (RNFL) thickness in primary open-angle glaucoma (POAG) patients and controls.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!