Background: The results of an in vivo study on the "ratio method" used in electronic foramen locators (EFL) are presented. EFLs are becoming widely used in the determination of the working length (WL) during the root canal treatment. The WL is the distance from a coronal reference point to the point at which canal preparation and filling should terminate. The "ratio method" was assessed by many clinicians with the aim of determining its ability to locate the apical foramen (AF). Nevertheless, in vivo studies to assess the method itself and to explain why the "ratio method" is able to locate the apical foramen and is unable to determine intermediate distances were not published so far.

Methods: A developed apparatus applies an electrical current signal with constant amplitude of 10 μARMS through the endodontic file within the root canal. The applied current signal is composed by summing six sine waves, from 250 Hz to 8 kHz. Data were acquired with the endodontic file tip at 7 different positions within root canals. In the frequency domain the quotients between the amplitude of a reference frequency and the amplitudes of the other frequencies components were calculated. Twenty one root canals were analyzed in vivo, during the endodontic treatment of twelve teeth of different patients, with age between 20 to 55 years.

Results: For the range of frequencies used in the commercial EFLs and for distances ranging from -3 mm to -1 mm of the AF, the impedance of the root canal is mainly resistive. However, when the file tip gets closer to AF, the root canal electrical impedance starts to change from a mainly resistive to a complex impedance. This change in the measured root canal impedance starts when the file tip is near -1.0 mm from the AF, getting stronger as the file tip gets closer to the AF. This change in the impedance behavior affects the ratio (quotient) of the impedance measured at different frequencies. Through graphic analysis it is demonstrated why EFLs based on the ratio method are unable to accurately measure any distances between - 3.0 and -0.5 mm from the apical foramen. The only reliable measurement is the 0 mm distance, which is when the file tip is at the AF.

Conclusions: The electrical impedance values of 21 root canals were in vivo studied. The results confirm the ability of EFLs that are based on the ratio method to accurately locate the AF position and explain why they are unable to determine the file tip position along the root canal.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2944265PMC
http://dx.doi.org/10.1186/1475-925X-9-46DOI Listing

Publication Analysis

Top Keywords

root canal
24
ratio method
12
"ratio method"
12
apical foramen
12
root canals
12
root
9
impedance
8
electronic foramen
8
foramen locators
8
locate apical
8

Similar Publications

Effect of different root canal irrigation regimes microbubble emulsion (MBE) via riboflavin photosensitizer (RFP), cerium oxide (CeO) nanoparticles (NPs), and Nd: YAP laser on antibacterial efficiency, microhardness (MH), smear layer (SL) removal efficacy, and push-out bond strength (PBS) of AH plus sealer to canal dentin. Sixty single-rooted teeth were selected, disinfected, and categorized into four groups based on the type of disinfection. Following disinfection, a pair of samples were randomly selected and visualized under scanning electron microscope (SEM) for SL evaluation.

View Article and Find Full Text PDF

Background: This study aims to investigate the prevalence and morphology of middle mesial canal (MMC) in mandibular first molar (M1M) among a Northwestern Chinese population, and to analyze their relationship with anatomical aspects of the mesial root.

Methods: Cone beam computed tomography (CBCT) was utilized to evaluate 898 M1Ms and assess the incidence and morphology of MMC. The following parameters for M1M with or without MMC were obtained: the vertical distance between the first appearance of MMC and canal orifices (D), the distance between mesiobuccal (MB) and mesiolingual (ML) canals (D1), the buccolingual width(L1) and mesiodistal width (L2) of mesial roots, and the flatness degree(L1/L2) of mesial roots.

View Article and Find Full Text PDF

Background: The smear layer formed during root canal instrumentation negatively affects root canal irrigation activity, which in turn can affect the treatment prognosis of endodontic treatment.

Aim: The aim of this study is to compare the efficiency of smear layer and debris removal in root canals using different irrigation protocols using scanning electron microscopy (SEM).

Materials And Methods: The quality of smear layer removal throughout the root canal was assessed in 30 intact extracted teeth divided into 3 groups according to the irrigation protocol: Group 1: 3% sodium hypochlorite (NaOCL) alternately with 17% ethylenediaminetetraacetic acetate (EDTA) was used.

View Article and Find Full Text PDF

Investigate the impact of antimicrobial photodynamic therapy (aPDT) using different photosensitizers (PSs) such as indocyanine green (IG), curcumin (CC), and methylene blue (MB), with or without intracanal application of calcium hydroxide (CH), on the push-out bond strength of glass-fiber posts (GFPs) to intraradicular dentin, the chemical composition of the root substrate, and the sealing of the adhesive interface across different thirds of intraradicular dentin. A total of 112 bovine teeth underwent biomechanical preparation and were divided into eight experimental groups (n = 14 each): Negative control with deionized water; positive control with deionized water + CH; IG group with indocyanine green and infrared laser; IG + CH group; CC group with curcumin and blue LED; CC + CH group; MB group with methylene blue and red laser; and MB + CH group. The push-out bond strength was measured using a universal testing machine (n = 8), and scanning electron microscopy characterized the fracture patterns.

View Article and Find Full Text PDF

Aim: To develop and validate an artificial intelligence (AI)-powered tool based on convolutional neural network (CNN) for automatic segmentation of root canals in single-rooted teeth using cone-beam computed tomography (CBCT).

Methodology: A total of 69 CBCT scans were retrospectively recruited from a hospital database and acquired from two devices with varying protocols. These scans were randomly assigned to the training (n = 31, 88 teeth), validation (n = 8, 15 teeth) and testing (n = 30, 120 teeth) sets.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!