Detection of bacterial biofilm on cochlear implants removed because of device failure, without evidence of infection.

Otol Neurotol

Department of Otorhinolaryngology, Head and Neck Surgery, Lariboisière Hospital, Assistance Publique des Hôpitaux de Paris & CNRS CESEM 8194, University Paris VII, Paris, France.

Published: October 2010

Objective: To investigate the formation of bacterial biofilms on the surface of the electrode array of cochlear implants (CI) explanted because of device failure, without evidence of infection, by use of scanning electron microscopy (SEM) and confocal laser scanning microscopy (CLSM).

Study Design: Prospective study.

Setting: Patients from 2 tertiary-care referral centers.

Patients And Methods: CIs were explanted from 9 patients because of device failure. Specimens were immediately snap-frozen in cold isopenthane, stored at -80°C and examined with SEM and CLSM by 3 investigators.

Main Outcome Measure: Presence of bacterial biofilm ascertained by SEM and CSLM.

Results: One specimen showed the formation of a bacterial biofilm on the middle ear part of the electrode array. No biofilm formation was found in the inner-ear part of electrode arrays. In the middle-ear part of the electrode array, a cylindrical cover of human muscular tissue was seen plugging the cochleostomy.

Conclusion: This is the first study demonstrating that bacterial biofilms may exist on the surface of the electrode array of CIs explanted because of device failure but not infection. We found 1 case of biofilm formation in 9 explanted CIs. Further studies with larger series of CIs are required to investigate biofilm formation on the surface of CI electrode arrays to address both the pathophysiology of bacterial biofilms and prevention of device-related infections in CI patients.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAO.0b013e3181e3d36bDOI Listing

Publication Analysis

Top Keywords

device failure
16
electrode array
16
bacterial biofilm
12
bacterial biofilms
12
surface electrode
12
biofilm formation
12
cochlear implants
8
failure evidence
8
evidence infection
8
formation bacterial
8

Similar Publications

Background: GLP-1 RAs improve cardiometabolic outcomes in obese, diabetic, and heart failure patients. Data on the safety and efficacy of GLP-1 RA in advanced heart failure with durable LVAD is limited.

Objectives: To assess the safety and efficacy of GLP-1 RA in durable LVAD patients.

View Article and Find Full Text PDF

Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD) is an alternative to traditional ICDs. The PRAETORIAN score, based on chest radiographs, has been validated to predict the probability of successful S-ICD defibrillation testing by assessing factors like fat thickness between the coil and sternum and generator placement.

Objective: This study evaluated the correlation between the PRAETORIAN score and clinical characteristics, as well as implantation variables.

View Article and Find Full Text PDF

Background: Vitamin E-diffused highly cross-linked polyethylene (HXLPE/Vit E) is a relatively advanced material used in total hip arthroplasty (THA) but whether it shows superiority is unclear.

Objective: This meta-analysis was performed to investigate the effect of HXLPE/Vit E liners in THA.

Methods: Medline/PubMed, Embase and Cochrane Library databases were searched to retrieve studies assessing the efficacy of HXLPE/Vit E liners in THA with the design of a randomized, controlled trial.

View Article and Find Full Text PDF

Worsening spasticity due to catheter breakage during intrathecal baclofen therapy: a case report.

J Med Case Rep

January 2025

Department of Orthopedic Surgery, Graduate School of Medicine, Kanazawa University, 13-1 Takara-machi, Kanazawa City, Ishikawa, 920-8641, Japan.

Background: Intrathecal baclofen therapy can substantially improve symptoms in patients with severe spasticity owing to traumatic spinal cord injury, multiple sclerosis, cerebral paresis, or tethered cord syndrome. Problems associated with intrathecal catheters include migration, laceration, occlusion, or disconnection. Several case reports have described the management of catheter fragments.

View Article and Find Full Text PDF

Background: In medial mobile-bearing unicompartmental knee arthroplasty (MB-UKA), the position of the bearing does not correspond to the planned position which will increasing the risk of bearing dislocation. This study aimed to explore the relationship between the malposition of the femoral and tibial components and the phenomenon of bearing deviation using postoperative radiological measurements.

Methods: One hundred twenty patients who underwent mobile-bearing uni-compartmental knee arthroplasty (MB-UKA) at our hospital between January and August 2023 were enrolled in this retrospective study.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!