Reference dose levels for dental panoramic radiography in Gwangju, South Korea.

Radiat Prot Dosimetry

Department of Oral and Maxillofacial Radiology, School of Dentistry, Dental Science Research Institute, Chonnam National University, 77 Yongbong-ro Buk-gu, Gwangju 500-757, South Korea.

Published: December 2010

This study assessed the reference dose levels for dental panoramic radiography in Gwangju city, South Korea based on the dose width product (DWP) and compared them with those already established elsewhere. A total of 44 panoramic dental radiographic sets (36 digital and 8 analogue panoramic sets) in 41 dental clinics in Gwangju city were chosen. The third quartile DWP was determined from 429 surface dose measurements of the adult surface dose in panoramic dental radiography. The third quartile DWP for panoramic radiography was 60.1 mGy mm. The proposed DWP reference levels of 60.1 mGy mm were less than or equal to those previously reported in other countries, such as Italy and UK, and acceptable for panoramic radiography in Gwangju, South Korea.

Download full-text PDF

Source
http://dx.doi.org/10.1093/rpd/ncq189DOI Listing