Shaping and timing gradient pulses to reduce MRI acoustic noise.

Magn Reson Med

University Medical Center Groningen, Department of Radiology, University of Groningen, Groningen, The Netherlands.

Published: August 2010

A method to reduce the acoustic noise generated by gradient systems in MRI has been recently proposed; such a method is based on the linear response theory. Since the physical cause of MRI acoustic noise is the time derivative of the gradient current, a common trapezoid current shape produces an acoustic gradient coil response mainly during the rising and falling edge. In the falling edge, the coil acoustic response presents a 180 degrees phase difference compared to the rising edge. Therefore, by varying the width of the trapezoid and keeping the ramps constant, it is possible to suppress one selected frequency and its higher harmonics. This value is matched to one of the prominent resonance frequencies of the gradient coil system. The idea of cancelling a single frequency is extended to a second frequency, using two successive trapezoid-shaped pulses presented at a selected interval. Overall sound pressure level reduction of 6 and 10 dB is found for the two trapezoid shapes and a single pulse shape, respectively. The acoustically optimized pulse shape proposed is additionally tested in a simulated echo planar imaging readout train, obtaining a sound pressure level reduction of 12 dB for the best case.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.22366DOI Listing

Publication Analysis

Top Keywords

acoustic noise
12
mri acoustic
8
gradient coil
8
falling edge
8
sound pressure
8
pressure level
8
level reduction
8
pulse shape
8
gradient
5
acoustic
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!