Variable classifications of glycemic index determined by glucose meters.

J Clin Biochem Nutr

Department of Food Science, National Pingtung University of Science and Technology, 1, Shuefu Road, Neipu, Pingtung 91201, Taiwan.

Published: July 2010

THE STUDY EVALUATED AND COMPARED THE DIFFERENCES OF GLUCOSE RESPONSES, INCREMENTAL AREA UNDER CURVE (IAUC), GLYCEMIC INDEX (GI) AND THE CLASSIFICATION OF GI VALUES BETWEEN MEASURED BY BIOCHEMICAL ANALYZER (FUJI AUTOMATIC BIOCHEMISTRY ANALYZER (FAA)) AND THREE GLUCOSE METERS: Accue Chek Advantage (AGM), BREEZE 2 (BGM), and Optimum Xceed (OGM). Ten healthy subjects were recruited for the study. The results showed OGM yield highest postprandial glucose responses of 119.6 +/- 1.5, followed by FAA, 118.4 +/- 1.2, BGM, 117.4 +/- 1.4 and AGM, 112.6 +/- 1.3 mg/dl respectively. FAA reached highest mean IAUC of 4156 +/- 208 mg x min/dl, followed by OGM (3835 +/- 270 mg x min/dl), BGM (3730 +/- 241 mg x min/dl) and AGM (3394 +/- 253 mg x min/dl). Among four methods, OGM produced highest mean GI value than FAA (87 +/- 5) than FAA, followed by BGM and AGM (77 +/- 1, 68 +/- 4 and 63 +/- 5, p<0.05). The results suggested that the AGM, BGM and OGM are more variable methods to determine IAUC, GI and rank GI value of food than FAA. The present result does not necessarily apply to other glucose meters. The performance of glucose meter to determine GI value of food should be evaluated and calibrated before use.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2901763PMC
http://dx.doi.org/10.3164/jcbn.10-08DOI Listing

Publication Analysis

Top Keywords

+/-
12
glucose meters
8
glucose responses
8
+/- faa
8
+/- +/-
8
faa
5
variable classifications
4
classifications glycemic
4
glycemic determined
4
glucose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!