Background: Hand involvement in scleroderma is a serious concern. Clinical tests to asses hand dysfunction are based on the experience of the clinician.

Objective: To asses if utrasonographic (US) measurement of A1 pulley thickness may be used as an indicator of hand mobility in scleroderma.

Materials And Methods: Institutional review board approval and patient informed consent was obtained. Twenty-eight patients affected suffering from scleroderma and 40 healthy controls were prospectively evaluated by two blinded radiologists with US, with a transducer operating at 17MHz. A1 pulley thickness was measured and correlated with the Hand Mobility in Scleroderma Test (HAMIS) and disease duration.

Results: The thickness of the A1 pulley was greater in sclerodermic patients than in controls (p < 0.05). Intra and inter-observer agreement were better for ultrasound (0.94;0.88) than for HAMIS tests (0.71;0.70). A good correlation between pulley thickness, hand mobility and disease duration was found (r = 0.78, p < 0.018; r = 0.54, p < 0.05).

Conclusion: A1 pulley thickness measured on ultrasound correlates with hand mobility and disease duration. Ultrasound is an useful tool to evaluate hand disability in scleroderma.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejrad.2010.05.013DOI Listing

Publication Analysis

Top Keywords

hand mobility
20
pulley thickness
16
hand
8
mobility scleroderma
8
thickness measured
8
mobility disease
8
disease duration
8
thickness
6
mobility
5
scleroderma
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!