Background: The use of the HeartMate II continuous-flow left ventricular assist device (LVAD) improves survival, quality of life and functional capacity of patients with advanced heart failure. However, no study so far has shown the benefits of these devices for cardiac function during peak exercise.

Objective: To assess cardiac and exercise performance in patients implanted with the HeartMate II LVAD under two settings: (i) optimal device support and (ii) reduced device support.

Methods: Twelve patients implanted with a HeartMate II LVAD performed a graded cardiopulmonary exercise test with respiratory gas exchange and non-invasive (rebreathing) haemodynamic measurements. After a 4 h resting period, patients performed an additional cardiopulmonary exercise test with reduced LVAD support (pump speed was reduced from optimal 9000-9600 to 6000 revs/min).

Results: In response to reduced HeartMate II LVAD support, resting cardiac power output and cardiac output decreased by 21% and 13%, respectively. Also at reduced device support, peak exercise cardiac power output was 39% lower (1.40+/-0.50 vs 2.31+/-0.58 W; p<0.05), peak cardiac output 30% lower (8.6+/-2.5 vs 12.2+/-2.1 l/min; p<0.05) and mean blood pressure 13% lower (74.3+/-14.9 vs 85.4+/-15.4 vs mm Hg; p<0.05). Exercise capacity was also diminished with 23% lower peak oxygen consumption (14.1+/-5.3 vs 18.2+/-4.5 ml/kg/min; p<0.05) and an 18% shorter exercise duration (516+/-119 vs 628+/-192 s; p<0.05).

Conclusion: It has been shown for the first time that the HeartMate II LVAD can confer both resting and peak cardiac functional benefits to patients with end-stage heart failure, thus improving exercise capacity.

Download full-text PDF

Source
http://dx.doi.org/10.1136/hrt.2010.193698DOI Listing

Publication Analysis

Top Keywords

device support
12
heartmate lvad
12
continuous-flow left
8
left ventricular
8
ventricular assist
8
assist device
8
cardiac exercise
8
exercise performance
8
patients implanted
8
implanted heartmate
8

Similar Publications

Left atrial appendage occlusion (LAAO) has become an important therapeutic target for stroke prevention in patients with nonvalvular atrial fibrillation. Over the past 2 decades, several advancements in LAAO devices (percutaneous and surgical) have been made for stroke prevention and arrhythmia therapy. However, there are several unanswered questions regarding optimal patient selection, the preferred LAAO approach and device, the management of periprocedural and postprocedural complications, including pericardial effusion, device-related thrombus, and device leaks.

View Article and Find Full Text PDF

Background: Growth differentiation factor (GDF)-15 is a pleiotropic cytokine that is associated with appetite-suppressing effects and weight loss in patients with malignancy.

Objectives: This study aims to investigate the relationships between GDF-15 levels, anorexia, cachexia, and clinical outcomes in patients with advanced heart failure with reduced ejection fraction (HFrEF).

Methods: In this observational, retrospective analysis, a total of 344 patients with advanced HFrEF (age 58 ± 10 years, 85% male, 67% NYHA functional class III), underwent clinical and echocardiographic examination, body composition evaluation by skinfolds and dual-energy x-ray absorptiometry, circulating metabolite assessment, Minnesota Living with Heart Failure Questionnaire, and right heart catheterization.

View Article and Find Full Text PDF

Background: Left ventricular (LV) dilatation and extensive scar portend a poor prognosis in heart failure (HF). The Revivent TC system (BioVentrix Inc) is used either during a hybrid transcatheter-surgical or a surgical-only procedure to exclude transmural scar and reduce LV dimensions.

Objectives: The purpose of this study was to examine the safety and efficacy of the Revivent TC® anchor system in patients with HF.

View Article and Find Full Text PDF

The PERFORMANCE II Trial: A Prospective Multicenter Investigation of a Novel Carotid Stent System.

JACC Cardiovasc Interv

December 2024

Department for Angiology, Brandenburg Medical School Theodor Fontane, Campus Clinic Brandenburg, Center for Internal Medicine I, Berlin, Germany; Department of Angiology, Sankt-Gertrauden-Krankenhaus, Berlin, Germany.

Background: Several randomized clinical trials have shown that the composite endpoint of death, stroke, and myocardial infarction (MI) is equivalent between carotid artery stenting and carotid endarterectomy. However, the risk of minor stroke has been consistently higher with carotid artery stenting.

Objectives: The authors sought to evaluate the safety and effectiveness of a novel carotid stent system comprised of a stent, an adjustable integrated embolic filter and a postdilation balloon, in patients at elevated risk for adverse events from carotid endarterectomy.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!