Objective: To assess the accuracy of measuring the cortical bone thickness adjacent to dental implants using two cone beam computed tomography (CBCT) systems.

Material And Methods: Ten 4 x 11 mm Astra Tech implants were placed at varying distances from the cortical bone in two prepared bovine ribs. Both ribs were scanned in a reproducible position using two different CBCT scanners. Ten examiners each carried out four measurements on all 10 implants using the two CBCT systems: vertical distance between the top of the implant and the alveolar crest (IT-AC), and thickness of the cortical bone from the outer surface of the implant threads at 3, 6 and 9 mm from the top of the implant. Ground sections were prepared and bone thickness was measured using a light microscope and a graticule to give a gold standard (GS) measurement.

Results: The examiner's measurements were significantly different between CBCT systems for the vertical and thickness dimensions (P<0.001) while measuring the cortical bone thickness between 0.3 and 3.7 mm. Within that range, i-CAT NG measurements were consistently underestimated in comparison with the GS. Accuitomo 3D60 FPD measurements closely approximated the GS, except when cortical bone thickness was <0.8 mm. The mean percentage errors from the GS at 3, 6 and 9 mm measurement levels were 68%, 28% and 18%, respectively, for i-CAT NG and 23%, 5% and 6%, respectively, for Accuitomo 3D60 FPD.

Conclusions: Within the limitations of this study, it was concluded that i-CAT NG (voxel size 0.3) may not produce sufficient resolution of the thin cortical bone adjacent to dental implants and, therefore, the measurements may not be accurate; whereas, Accuitomo 3D60 FPD (voxel size 0.125) may produce better resolution and more accurate measurement of the thin bone.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1600-0501.2009.01905.xDOI Listing

Publication Analysis

Top Keywords

cortical bone
16
bone thickness
12
accuracy measuring
8
measuring cortical
8
thickness adjacent
8
adjacent dental
8
dental implants
8
implants cone
8
cone beam
8
beam computed
8

Similar Publications

Comparative evaluation of allograft particulate bone and cortical bone blocks combined with xenograft bone for labial bone defects in the aesthetic zone: a prospective cohort study.

BMC Oral Health

January 2025

State Key Laboratory of Oral Diseases & National Clinical Research Center for Oral Diseases & Department of Implant Dentistry, West China Hospital of Stomatology, Sichuan University, Chengdu, 610041, China.

Purpose: This study aimed to evaluate the osteogenic performance of allograft particulate bone and cortical bone blocks combined with xenograft under bovine pericardium membranes, for treating different degrees of labial bone defects in the aesthetic zone.

Materials And Methods: Twenty-four patients with bone defects were divided into two groups based on defect severity (Terheyden 1/4 and 2/4 groups). The Terheyden 1/4 group received granular bone grafts alone, while the Terheyden 2/4 group received cortical bone blocks combined with granular bone grafts.

View Article and Find Full Text PDF

HIV-related mortality has fallen due to scale-up of antiretroviral therapy (ART), so more women living with HIV (WLH) now live to reach menopause. Menopausal estrogen loss causes bone loss, as do HIV and certain ART regimens. However, quantitative bone data from WLH are few in Africa.

View Article and Find Full Text PDF

Osteoporosis is the most common bone metabolic unbalance, leading to fragility fractures, which are known to be associated with structural changes in the bone. Cortical bone accounts for 80 % of the skeleton mass and undergoes remodeling throughout life, leading to changes in its thickness and microstructure. Although many studies quantified the different cortical bone structures using CT techniques (3D), they are often realised on a small number of samples.

View Article and Find Full Text PDF

Determinants for successful medullary fixation of the superior ramus.

Injury

January 2025

Department of Orthopaedic Surgery, Cedars - Sinai Medical Center, Los Angeles, CA, USA. Electronic address:

Objectives: The purpose of this study is to determine what demographic and anatomical variables affect successful placement of a superior medullary ramus screw, and how they affect the maximal diameter of that screw.

Methods: Design: Prognostic Level IV SETTING: Level I Trauma Center Patients/Participants: Two hundred consecutive patients underwent computed tomography (CT) of the pelvis. We included those patients aged 18 and older without osseous injury or abnormalities precluding measurement.

View Article and Find Full Text PDF

Background: This is a novel rat study using native peptide therapy, focused on reversing quadriceps muscle-to-bone detachment to reattachment and stable gastric pentadecapeptide BPC 157 per-oral therapy for shared muscle healing and function restoration.

Methods: Pharmacotherapy recovering various muscle, tendon, ligament, and bone lesions, and severed junctions (i.e.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!