Monte Carlo calculation of dosimetry parameters for the IR08-103Pd brachytherapy source.

Med Phys

Engineering Faculty, Research and Science Campus, Islamic Azad University, 1477893855 Tehran, Iran.

Published: June 2010

Purpose: For the treatment of some cancerous tumors using brachytherapy methods and low-energy photon sources, such as 125I and 103Pd, the American Association of Physicists in Medicine Task Group No. 43U1 report recommends that the dosimetric parameters of a new brachytherapy source must be determined in two experimental and Monte Carlo theoretical methods before using each new source clinically. This study presents the results of Monte Carlo calculations of the dosimetric parameters for IR08-103Pd brachytherapy source design. IR08-103Pd seed has been manufactured at the Agricultural, Medical and Industrial Research School.

Methods: Version 5 of the (MCNP) Monte Carlo radiation transport code was used to calculate the dosimetry parameters around the source. Three geometric models of the seed, based on different locations of beads inside the titanium capsule, were simulated. The seed contains five resin beads of 0.6 mm diameter having 103Pd uniformly absorbed in the bead volume, which were contained within a cylindrical titanium capsule having 0.8 mm outside diameter and 4.8 mm length.

Results: The Monte Carlo calculated dose rate constant of the IR08-103Pd seed was found to be 0.695 +/- 0.021 cGyU(-1) h(-1). Also in this study, the geometry function G(r, theta), line and point-source radial dose functions gL(r) and gP(r), and the anisotropy function F(r, theta), have been calculated at distances from 0.25 to 7 cm. The results of these calculations have been compared with measured values for an actual IR08-103Pd seed.

Conclusions: There are no statistical significant dosimetric differences among the three seed orientations in this study (i.e., ideal, vertical, and diagonal). However, the observed differences between the calculated and measured values could be explained by the measurement uncertainty and the configuration of the resin beads within the capsule and capsule orientation.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.3416922DOI Listing

Publication Analysis

Top Keywords

monte carlo
20
brachytherapy source
12
dosimetry parameters
8
parameters ir08-103pd
8
ir08-103pd brachytherapy
8
dosimetric parameters
8
ir08-103pd seed
8
titanium capsule
8
resin beads
8
function theta
8

Similar Publications

Objectives: To assess the pharmacokinetics and pharmacodynamics of imipenem in a retrospective cohort of hospitalized Chinese older patients.

Methods: A population pharmacokinetic (PPK) model was constructed utilizing a nonlinear mixed-effects modeling approach. The final model underwent evaluation through bootstrap resampling and visual predictive checks.

View Article and Find Full Text PDF

Purpose In linac-based stereotactic radiosurgery (SRS) utilizing a multileaf collimator (MLC) for brain metastases (BMs), a volumetric-modulated arc (VMA) technique is indispensable for generating a suitable dose distribution with efficient planning and delivery. However, the optimal calculation grid spacing (GS) and statistical uncertainty (SU) of the Monte Carlo algorithm for VMA optimization have yet to be determined. This planning study aimed to examine the impacts of GS and GU settings on VMA-based SRS planning and to find the optimal combination for templating.

View Article and Find Full Text PDF

Background: Uncertainty about optimal tranexamic acid (TXA) dosage has led to significant practice variation in hip arthroplasty. We aimed to identify the optimal i.v.

View Article and Find Full Text PDF

Importance: Patients with uterine factor infertility (UFI) have few options for family building. Uterus transplant is a feasible treatment for some patients; however, cost remains a significant concern.

Objective: To compare the cost effectiveness of treatment for patients with absolute uterine factor infertility to achieve 1-2 singleton births by gestational carrier or uterus transplant DESIGN: Decision analysis from the United States healthcare sector perspective, with time horizons to achieve one or two singleton births.

View Article and Find Full Text PDF

Heterogeneous head phantom for validating treatment planning system in boron neutron capture therapy.

Appl Radiat Isot

January 2025

Institute of Nuclear Engineering and Science, National Tsing Hua University, 101, Sec. 2, Kuang-Fu Road, Hsinchu, 30013, Taiwan; Nuclear Science and Technology Development Center, National Tsing Hua University, 101, Sec. 2, Kuang-Fu Road, Hsinchu, 30013, Taiwan. Electronic address:

In clinical boron neutron capture therapy (BNCT), the distribution of dose to a heterogeneous medium that is predicted by a treatment planning system (TPS) should be experimentally validated. A head phantom specifically developed for this purpose is described and demonstrated herein. The cylindrical phantom exhibits distinct regions made from four materials (polymethyl methacrylate, calcium phosphate, air, and boric acid) to approximate a head structure with explicitly defined skin, skull, and brain tissue with a cavity and tumor within.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!