Purpose: The aim of this study was to evaluate the effect of different levels of unilateral angular misfit on preload maintenance of retention screws of single implant-supported prostheses submitted to mechanical cycling.
Materials And Methods: Premachined UCLA abutments were cast with cobalt-chromium alloy to obtain 48 crowns divided into four groups (n=12). The crowns presented no misfit in Group A (control group) and unilateral misfits of 50 μm, 100 μm and 200 μm in the groups B, C and D, respectively. The crowns were attached to external hexagon implants with a titanium retention screw with torque of 30 N/cm. Oblique loading of 130 N at 2 Hz was applied on each replica, totalizing 5×10⁴ and 1×10⁶ cycles. Detorque values were measured initially and after each cycling period. Data were evaluated by analysis of variance and Tukey's HSD test (p<0.05).
Results: All groups presented reduced initial detorque values (p<0.05) in comparison to the insertion torque (30±0.5 N/cm) and Group A (25.18 N/cm) exhibited the lowest reduction. After mechanical cycling, all groups presented detorque values from 19.5 N/cm to 22.38 N/cm and the mechanical cycling did not statistically influence the detorque values regardless the misfit level of the replicas.
Conclusion: The unilateral misfit influenced the preload maintenance only before mechanical cycling. The mechanical cycling did not influence the torque reduction.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jpor.2010.05.002 | DOI Listing |
Materials (Basel)
July 2023
Department of Reconstructive Oral Care, Academic Centre for Dentistry Amsterdam (ACTA), Universiteit van Amsterdam and Vrije Universiteit, 1081 LA Amsterdam, The Netherlands.
This study investigated the effects of antirotational titanium bases on the mechanical behavior of CAD/CAM titanium bases used for implant-supported prostheses. The aim was to assess the impact on the marginal fit, fatigue behavior, stress concentration, and fracture load of implant-supported CAD/CAM zirconia crowns. Forty titanium implants were divided into two groups: those with antirotational titanium bases (ARs) and those with rotational titanium bases (RTs).
View Article and Find Full Text PDFTo evaluate hard and soft tissue response and mechanical integrity for screw-retained layered zirconia crowns bonded to titanium nitride-coated titanium (TiN) CAD/CAM abutments supported by implants at 1- and 2-year follow-ups. A total of 46 patients were restored with a total of 102 free-standing implant-supported layered zirconia crowns, which were bonded to their respective abutments in the dental laboratory and delivered as a one-piece screw-retained crown. Baseline, 1-year, and 2-year data were collated on pocket probing depth, bleeding on probing, marginal bone levels, and mechanical complications.
View Article and Find Full Text PDFJ Oral Sci
October 2022
Department of Prosthodontics, Showa University School of Dentistry.
The purpose of this review is to search for complications of dental implant superstructures and consider the issues involved. This narrative review was performed by searching through PubMed databases and review articles that were published after 1990. Misfitting of the superstructure can result in loosening of screws, reduced preload, and in some cases, significant stress around the implant.
View Article and Find Full Text PDFJ Adv Prosthodont
February 2022
Department of Prosthetic Dental Sciences, College of Dentistry, King Saud University, Riyadh, Kingdom of Saudi Arabia.
Purpose: The purpose of this study was to investigate the misfit and screw preload at the implant abutment connection of implant supported fixed dental prosthesis with cantilever (ICFDP) manufactured using different digital manufacturing techniques and to compare the screw preload before and after cyclic loading.
Materials And Methods: Mandibular jaw model with four intra-foraminal implants was scanned using digital scanner. Stereolithography file was used to design a framework with nonengaging (NE) abutments and 10 mm cantilever distal to one terminal implant.
Int J Environ Res Public Health
November 2020
Porto Dental Institute, 4150-518 Porto, Portugal.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!