Background: The purpose of this study was to compare the stability of a 2.4mm palmar locking compression plate and a new intramedullary nail-plate-hybrid Targon DR for dorsally comminuted distal radius fractures.

Methods: An extraarticular 10mm dorsally open wedge osteotomy was created in 8 pairs of fresh frozen human radii to simulate an AO-A3-fracture. The fractures were stabilized using one of the fixation methods. The specimens were loaded axially with 200 N and dorsal-excentrically with 80 N. 2000cycles of dynamic loading and axial loading-to-failure were performed.

Findings: Axial loading revealed that intramedullary osteosynthesis (Targon DR: 369 N/mm) was significantly (p=0.017) stiffer than plate osteosynthesis (Locking compression plate: 131 N/mm). With 214 N/mm the intramedullary nail also showed higher stability during dorsal excentric loading than the Locking compression plate with 51 N/mm (p=0.012). After 2000 cycles of axial loading with 80 N the Targon DR-group was significantly stiffer than the Locking compression plate-group under both loading patterns. Neither group showed significant changes in stiffness after 2000 cycles. Under dorsal excentric loading the Targon DR-group was still significantly stiffer with 212 N/mm than the Locking compression plate-group with 45 N/mm (p=0.012). The load to failure tests demonstrated higher stability of intramedullary nailing (625 N) when compared to plate osteosynthesis (403 N) (p<0.025).

Interpretation: The study shows that intramedullary fixation of a distal AO-A3 radial fracture is biomechanically more stable than volar fixed-angle plating under axial and dorsal-excentric loading in an experimental setup.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiomech.2010.06.004DOI Listing

Publication Analysis

Top Keywords

locking compression
20
compression plate
12
intramedullary nailing
8
dorsally comminuted
8
comminuted distal
8
distal radius
8
axial loading
8
plate osteosynthesis
8
higher stability
8
dorsal excentric
8

Similar Publications

Locking plates have a rapidly growing process especially in the past decades and results are satisfactory especially in the osteoporotic bones compared to non-locking compression plates. There are many forms of failure in the fracture fixation of locking plates, and screw pull-out is one of the main failure reasons. In this study, we aim to investigate pull-out failure in locking plates using locking spongious screws.

View Article and Find Full Text PDF

Objective:  To determine the effect of locking head inserts (LHI) on plate strain, stiffness, and deformation when applied to a 3.5-mm broad locking compression plate (LCP) in an open fracture-gap model.

Study Design:  Six, 13-hole, 3.

View Article and Find Full Text PDF

Mandibular Reconstruction With a Patient-Specific Implant Following Surgical Excision of an Acanthomatous Ameloblastoma in a Dog.

J Vet Dent

January 2025

Department of Dentistry, Oral and Maxillo-facial Surgery, Eastcott Veterinary Referrals, Part of Linnaeus Group, Swindon, UK.

Canine acanthomatous ameloblastoma (CAA) is an invasive benign epithelial odontogenic tumour most commonly affecting the mandible of large breed dogs. To the author's knowledge, this report describes the first computer-aided design patient-specific implant (PSI) that has been placed for a critical sized bone defect in mandibular reconstruction of a dog in the UK. The aim was to restore mandibular stability using a regenerative approach combining a titanium locking plate and compression-resistant matrix infused with recombinant human bone morphogenetic protein-2 (rhBMP-2) to bridge the 85 mm mandibular defect created by a segmental mandibulectomy.

View Article and Find Full Text PDF

Background: Length unstable femoral shaft fractures in school-aged children usually demand surgical treatment, but the optimal choice remains controversial, especially in overweight adolescents. This study aimed to compare the clinical results of locking compression plates (LCP) and elastic stable intramedullary nails (ESIN) combined with temporary external fixator (TEF) in school-aged children weighing over 50 kg.

Methods: Between January 2010 and January 2018, children over 50 kg with length unstable femoral shaft fracture treated with ESIN & EF in the authors' institute were included in this study.

View Article and Find Full Text PDF

Treatment of posterolateral tibial plateau fractures through the articular line approach.

BMC Musculoskelet Disord

January 2025

Department of Orthopedics, Lishui Central Hospital, Zhejiang, 323000, China.

Background: Posterolateral tibial plateau fractures pose significant challenges for orthopedic surgeons due to the anatomical risks associated with the posterolateral approach. Despite numerous surgical techniques available, there lacks a consensus on the optimal approach.

Methods: Articular line incision approach was employed on 12 patients suffering from posterolateral tibial plateau fractures.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!