Background: In-vitro mechanical tests are commonly performed to assess pre-clinically the effect of implant design on the stability of hip endoprostheses. There is no standard protocol for these tests, and the forces applied vary between studies. This study examines the effect of the abductor force with and without application of the anterior-posterior hip contact force in the in-vitro assessment of cementless hip implant stability.

Methods: Cementless stems (VerSys Fiber Metal) were implanted in twelve composite femurs which were divided into two groups: group 1 (N = 6) was loaded with the hip contact force only, whereas group 2 (N = 6) was additionally subjected to an abductor force. Both groups were subjected to the same cranial-caudal hip contact force component, 2.3 times body weight (BW) and each specimen was subjected to three levels of anterior-posterior hip contact load: 0, -0.1 to 0.3 BW (walking), and -0.1 to 0.6 BW (stair climbing). The implant migration and micromotion relative to the femur was measured using a custom-built system comprised of 6 LVDT sensors.

Results: Substantially higher implant motion was observed when the anterior-posterior force was 0.6BW compared to the lower anterior-posterior load levels, particularly distally and in retroversion. The abductor load had little effect on implant motion when simulating walking, but resulted in significantly less motion than the hip contact force alone when simulating stair climbing.

Conclusions: The anterior-posterior component of the hip contact load has a significant effect on the axial motion of the stem relative to the bone. Inclusion of the abductor force had a stabilizing effect on the implant motion when simulating stair climbing.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2907319PMC
http://dx.doi.org/10.1186/1749-799X-5-40DOI Listing

Publication Analysis

Top Keywords

hip contact
28
contact force
16
anterior-posterior hip
12
contact load
12
abductor force
12
implant motion
12
hip
10
cementless hip
8
force
8
stair climbing
8

Similar Publications

Background: Few studies have assessed vertical ground reaction force, contact time, contact area, and center of pressure during gait in the early phase post-total hip arthroplasty. This study aimed to investigate whether these parameters are more pronounced in participants post-total hip arthroplasty compared to healthy controls.

Methods: We included 22 female participants who underwent total hip arthroplasty (age, 68.

View Article and Find Full Text PDF

Objective: To summarize the characteristics of posterior coronal banana-shaped fragments and its research progress in treatment of intertrochanteric femoral fracture with cephalomedullary nail, provide valuable reference for clinical practice.

Methods: Relevant domestic and foreign literature was extensively reviewed to summarize the history, anatomical structure of posterior coronal banana-shaped fragments, and the need for reduction and fixation, the influence on the stability of cephalomedullary nail, and the remedies.

Results: The posterior coronal banana-shaped fragments refers to the second level fracture line of the intertrochanteric femoral fracture, involving four anatomical structures, namely, the posterior part of the greater trochanter, the intertrochanteric crest, the lesser trochanter, and the posteromedial cortex.

View Article and Find Full Text PDF

Kinematical and Physiological Responses of Overground Running Gait Pattern at Different Intensities.

Sensors (Basel)

November 2024

Centre of Research, Education, Innovation and Intervention in Sport and Porto Biomechanics Laboratory, Faculty of Sport, University of Porto, 4200-450 Porto, Portugal.

Runners achieve forward locomotion through diverse techniques. However, understanding the behavior of the involved kinematical variables remains incomplete, particularly when running overground and along an intensity spectrum. We aimed to characterize the biomechanical and physiological adaptations while running at low, moderate, heavy and severe intensities.

View Article and Find Full Text PDF

Stem subsidence in total hip arthroplasty: retrospective investigation of a short stem using a simple measurement approach.

Arch Orthop Trauma Surg

December 2024

Department of General Orthopaedics and Tumour Orthopedics, Muenster University Hospital, Albert-Schweitzer-Campus 1, 48149, Muenster, Germany.

Background: Uncemented total hip arthroplasty (THA) is a successful treatment for advanced hip joint diseases. More recently, short stems became increasingly popular, but stem subsidence remains a concern. This study investigates early short stem subsidence in a large patient cohort using a simple measurement approach for everyday practice.

View Article and Find Full Text PDF

Understanding how forces are transmitted through the knee after TKA is essential, as it may explain why many patients experience pain or functional limitations during various activities. This study compared knee muscle forces and knee contact forces (KCF) during sit-to-stand in patients one year after unilateral total knee arthroplasty (TKA) with either a medial ball-and-socket (MBS) or posterior stabilized (PS) implant and compared them to a group of similarly healthy aged controls (CTRL). A musculoskeletal model and static optimization estimated lower limb kinematics, knee kinetics, muscle forces, and KCFs.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!