[Using HPLC-MS/MS method to analyze monohydroxy metabolites of PAHs in urine].

Wei Sheng Yan Jiu

National Institute for Occupational Health and Poison Control, Chinese Center for Disease Control and Prevention, Beijing 100050, China.

Published: May 2010

Objective: To establish a HPLC-MS/MS method for the simultaneous determination of 4 monohydroxy metabolites of polycyclic aromatic hydrocarbons (PAHs) in urine, including 1-naphthol, 2-naphthol, 3-phenanthrol and 1-pyrenol).

Methods: After being treated by enzymatic hydrolysis, urine samples were purified by liquid-liquid extraction, evaporated under nitrogen to constant volume and examined by HPLC-MS/MS. Target compounds were quantified by internal standard calibration curve.

Results: The reproducibility and precision were improved by internal standard calibration. The recovery of four target compounds were in the range of 89.7%-103.0%. The precision of intra- and inter-batch was 1.0%-4.2% and 1.8%-5.8% respectively. The limits of quantification was 0.2 microg/L for 1-naphthol, 2-naphthol and 1-pyrenol and the limits of quantification was 0.1 microg/L for 3-phenanthrol.

Conclusion: The sensitivity, reproducibility and recovery of HPLC-MS/MS method were good for the analysis of 1-naphthol, 2-naphthol, 3-phenanthrol and 1-pyrenol in the urine of people on occupational and nonoccupational exposure to PAHs.

Download full-text PDF

Source

Publication Analysis

Top Keywords

hplc-ms/ms method
12
1-naphthol 2-naphthol
12
monohydroxy metabolites
8
2-naphthol 3-phenanthrol
8
target compounds
8
internal standard
8
standard calibration
8
limits quantification
8
quantification microg/l
8
[using hplc-ms/ms
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!