Purpose: To report a retrospective evaluation of the 6-F Angio-Seal closure device in antegrade and retrograde common femoral artery (CFA) punctures during endovascular procedures in diabetic patients with critical limb ischemia (CLI).
Methods: From January 2005 to March 2009, 2374 diabetic CLI patients underwent interventional procedures in the lower limbs at a single center under systemic anticoagulation (heparin 70 U/kg). In this population, 2016 patients (1184 men; mean age 69.6+/-9.1 years) had 2372 CFA punctures treated with either manual compression [205 punctures in 161 (8.0%) patients] or Angio-Seal deployment (2167 punctures in 1855 patients) and were eligible for this analysis. In the study cohort, there were 1889 antegrade CFA punctures closed with the device in 1626 (87.6%) patients compared to 278 retrograde punctures sealed in 229 (12.4%) patients. The complications from the antegrade CFA punctures were compared to those from retrograde closure and manual compression.
Results: The success rate for achieving hemostasis after antegrade and retrograde Angio-Seal placement was 97.9% and 97.8%, respectively. Major complications following antegrade Angio-Seal deployment, retrograde Angio-Seal deployment, and manual compression occurred in 20/1889 (1.1%), 5/278 (1.8%), and 4/205 (2.0%) cases, respectively. All complications developed within 24 hours of the procedure. No further complications were recorded in the 18-month follow-up (range 1-36). The overall complication rates after antegrade puncture closure, retrograde puncture closure, and manual compression at 30 days was 2.5%, 4.0%, and 4.9%, respectively (p = NS).
Conclusion: This retrospective study shows that the 6-F Angio-Seal is a valuable and safe vascular closure device for percutaneous transfemoral antegrade access in diabetic patients undergoing interventional procedures for CLI.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1583/09-2960.1 | DOI Listing |
Ann Vasc Surg
December 2024
Department of Vascular Surgery, IRCCS Sacro Cuore-Don Calabria, Negrar, VR, Italy.
Background: ProGlide is a suture-mediated vascular closure device (VCD) indicated for retrograde access closure at the common femoral artery (CFA). However, its off-label use for antegrade and/or superficial femoral artery (SFA) access has become common in many practices. This study evaluated the efficacy and safety of ProGlide for femoral artery access closure in patients undergoing antegrade infrainguinal endovascular procedures.
View Article and Find Full Text PDFAnn Vasc Surg
January 2025
Department of Surgery and Science, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan.
Background: The treatment outcomes of vein and bovine pericardial patches in thromboendarterectomy (TEA) for common femoral artery (CFA) lesions in patients with peripheral arterial disease (PAD) remain unclear. The purpose of this study was to evaluate the efficacy and safety of the bovine pericardial patch by comparing it with vein patch angioplasty.
Methods: This was a multicenter retrospective study.
BMJ Case Rep
October 2024
Department of Neurosurgery, The Jikei University School of Medicine, Tokyo, Japan.
Vascular closure devices (VCDs) show fewer complications related to the puncture site than manual compression but can cause stenosis or occlusion of the common femoral artery (CFA). A patient in her 30s who underwent suture-mediated VCD for haemostasis at the right CFA puncture site after neurointervention showed occlusion of the right CFA on postoperative day 2. Endovascular treatment retrieved the thrombus from the occlusion site, and surgical removal of a suture causing stenosis between the dissected posterior wall intima and anterior wall allowed the resumption of full flow through the right CFA.
View Article and Find Full Text PDFCVIR Endovasc
October 2024
Department of Cardiovascular Medicine, Asahi General Hospital, I-1326 Asahi, Chiba, 289-2511, Japan.
Background: Hemostatic devices are now frequently used in femoral artery punctures, and the Angio-Seal (Terumo, Tokyo, Japan) is one of the most commonly used devices for closure of the femoral artery because it provides rapid hemostasis. Although device failure rarely occurs, if the collagen falls into the femoral artery, it may lead to severe limb ischemia. Herein, we describe a case of a novel endovascular technique for the treatment of Angio-Seal arterial closure device failure.
View Article and Find Full Text PDFJ Cardiol Cases
May 2024
The Second Department of Internal Medicine, University of Occupational and Environmental Health, Kitakyusyu, Japan.
Unlabelled: We aimed to describe a technique for approaching the common femoral artery (CFA) in cases where doing so is difficult owing to an occluded lesion caused by a previously implanted stent. A 72-year-old woman had severe stenotic lesions in both iliac arteries that required an approach via the bilateral femoral arteries. The right CFA had a previously implanted stent and a completely occluded lesion that extended from the superficial femoral artery (SFA).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!