Purpose: It is important to accurately measure intraocular pressure (IOP) in eyes with corneal endothelial dysfunction both before and after Descemet stripping with automated endothelial keratoplasty (DSAEK). Glaucoma is a common comorbidity in this population, and IOP elevation can worsen corneal edema. Additionally, preexisting glaucoma and steroid-responsive ocular hypertension are significant risk factors for graft rejection after DSAEK. Accurate tonometry is limited by variations in central corneal thickness (CCT) and corneal hydration that may affect corneal biomechanical properties. We analyzed CCT and IOP in eyes before and after DSAEK to determine whether changes in corneal biomechanics because of edema, grafted tissue, and subsequent stromal deturgescence affect IOP measurement.

Methods: A retrospective chart review was performed on 32 eyes from 31 patients with corneal edema secondary to Fuchs endothelial dystrophy, bullous keratopathy, or prior graft failure, or rejection that received uncomplicated DSAEK with no evidence of persistent corneal edema or steroid-induced ocular hypertension. IOP was measured by Tono-Pen XL, and CCT was measured by ultrasound pachymetry before and approximately 3 months after surgery. We used paired t tests to evaluate changes in CCT and IOP after DSAEK and linear regression to determine the relationship between CCT and IOP before and after surgery.

Results: CCT significantly decreased from 703 ± 82 to 650 ± 52 μm after DSAEK (P = 0.0026), but there was no significant change in measured IOP (16.7 ± 3.4 mm Hg preoperatively and 16.3 ± 4.1 mm Hg postoperatively; P = 0.61). There was no significant relationship between CCT and IOP before (slope = 0.10 ± 0.07 mm Hg/10 μm; r = 0.062; P = 0.17) or after (slope = 0.21 ± 0.14 mm Hg/10 μm; r = 0.072; P = 0.14) DSAEK.

Conclusion: CCT is significantly reduced by DSAEK but remains well above the normal range. IOP remains near the preoperative level 3 months after DSAEK. Furthermore, no correction is required for Tono-Pen measurements of IOP in corneas thickened by edema secondary to endothelial dysfunction or by DSAEK.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ICO.0b013e3181d25cbdDOI Listing

Publication Analysis

Top Keywords

cct iop
16
corneal edema
12
iop
11
corneal
9
dsaek
9
corneal thickness
8
intraocular pressure
8
descemet stripping
8
stripping automated
8
automated endothelial
8

Similar Publications

Purpose: This study aims to perform a clinical investigation of an innovative rebound technology-based device, the M-TONX, to simultaneously measure intraocular pressure (IOP) and central corneal thickness (CCT).

Methods: The IOP and CCT of the patients were first measured by the M-TONX. Then, the measurements were repeated by the Goldman applanation (GAT) and the Pentacam corneal topographer, as the standard devices.

View Article and Find Full Text PDF

American alligators () are an important apex predator of semiaquatic habitats of the southern United States. Commercial alligator farming has grown in the last several decades, leading to a need to understand the health conditions that affect this species to ensure appropriate management and welfare. The aims of this study were 1) to establish and describe normative data including results of common ophthalmic diagnostic tests and conjunctival flora, 2) document ocular pathology in this population, and 3) assess the effect of topical rocuronium bromide and 10% phenylephrine for facilitating pharmacologic mydriasis.

View Article and Find Full Text PDF

Intraocular pressure (IOP) readings using three different methods (Goldmann applanation tonometry (GAT), Corvis ST, and iCare) were compared in patients who underwent penetrating keratoplasty (PK). : An observational cross-sectional study with prospective recruitment of patients was conducted. IOP measurements were acquired using GAT, iCare, and Corvis (including both uncorrected IOP (CVS-IOP) and biomechanical IOP (bIOP)), and the agreement among methods was analyzed using Bland-Altman plots.

View Article and Find Full Text PDF

: This study aimed to describe the outcomes of a staged procedure combining Descemet membrane endothelial keratoplasty (DMEK) and sutureless scleral fixation (SSF) of a one-piece intraocular lens (IOL) in a case series. Co-performing endothelial keratoplasty (EK) and SSF is associated with intraoperative and postoperative complications such as graft deployment difficulties, air migration, graft detachment, and IOL opacification or tilt, all of which are evaluated in this study. This is a retrospective observational case series.

View Article and Find Full Text PDF

Purpose: To evaluate the efficiency of a hybrid tip in removing cataract grade 3 and 4 in high and low intraocular pressure (IOP) settings.

Methods: This was a randomized, prospective, double-arm contralateral eye study. One randomized eye of each subject had phacoemulsification with high IOP settings (50 mmHg) while the other eye had phacoemulsification with low IOP settings (20 mmHg).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!