Computational method to predict esophageal temperature elevations during pulmonary vein isolation.

Pacing Clin Electrophysiol

Al-Sabah Arrhythmia Institute and Division of Cardiology, St. Luke's and Roosevelt Hospitals, Columbia University College of Physicians & Surgeons, 1111 Amsterdam Avenue, New York, NY 10025, USA.

Published: October 2010

Background: The esophagus is in close proximity to the posterior wall of the left atrium, which renders it susceptible to thermal injury during radiofrequency (RF) ablation procedures for atrial fibrillation (AF). Real-time assessment of esophageal position and temperature (T°) during pulmonary vein (PV) isolation has not been extensively explored.

Objective: To develop a protocol that allows estimation of the potential for, and avoidance of, esophageal heating.

Methods: In consecutive patients who underwent PV isolation, a thermal probe was used to monitor T° fluctuations in the esophagus during application of RF energy. The tip of the thermal probe was positioned at the level of the targeted PV and RF was discontinued for T° rise >0.5°C. The proximity of individual PVs to the esophagus was measured from the temperature probe tip to the closest posterior part of the Lasso catheter from review of biplane projections (left anterior oblique 60° and right anterior oblique 30°). These raw distances were entered into the Pythagorean theorem and the actual distance between the esophageal thermal probe and PV antrum was determined.

Results: The study cohort included 44 patients (60 ± 11 years, 61% male, 57% lone AF). The thermal probe in the esophagus was closer to the left-sided PVs (left common pulmonary vein: 20.9 ± 13 mm, left upper pulmonary vein: 20.5 ± 11 mm, left lower pulmonary vein: 23.4 ± 10 mm) than the right-sided ones (right common pulmonary vein: 31.0 ± 11 mm, right upper pulmonary vein: 41.9 ± 18 mm, right lower pulmonary vein: 34.5 ± 16 mm). A T° increase >0.5° C occurred during 116/1,495 (7.8%) deliveries. A T° rise was more likely during ablation of left-sided PVs than right-sided PVs (55% vs 10%, P < 0.0001) and when RF was delivered ≤ 24 mm from the esophagus (sensitivity 91%, specificity 81%, positive predictive value 75%, and negative predictive value 93%).

Conclusion: A thermal probe placed in the esophagus provides real-time T° monitoring and anatomic localization. A T° rise is more likely during ablation of left PVs and during RF deliveries within 24 mm of the esophageal thermal probe.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.2010.02804.xDOI Listing

Publication Analysis

Top Keywords

pulmonary vein
32
thermal probe
24
t° rise
12
pulmonary
8
vein
8
vein isolation
8
anterior oblique
8
esophageal thermal
8
probe esophagus
8
left-sided pvs
8

Similar Publications

We present two patients with a history of paroxysmal atrial fibrillation who developed pulmonary vein stenosis (PVS) following atrial fibrillation (AF) ablation. Case 1 involved a female patient in her 50s who was asymptomatic for pulmonary symptoms but was found to have a high degree of left superior PVS 15 months after AF ablation. This was demonstrated using contrast-enhanced computed tomography (CE-CT) and supported by findings of perfusion defects on ventilation-perfusion (V/Q) scan.

View Article and Find Full Text PDF

Incidence, Characteristics and Prognostic Significance of Early Recurrences Following Different Ablation Approaches for Persistent Atrial Fibrillation.

Heart Rhythm

December 2024

University of Melbourne, Melbourne, Victoria, Australia; Royal Melbourne Hospital, Melbourne, Victoria, Australia; Baker Heart and Diabetes Research Institute, Melbourne, Victoria, Australia. Electronic address:

Background: Early recurrence of atrial tachyarrhythmia (ERAT) is common following catheter ablation of atrial fibrillation (AF). The specific clinical and arrhythmia characteristics of ERAT influencing late recurrence risk in persistent AF is unclear. Additionally, the impact of different ablation strategies on the incidence and prognostic significance of ERAT remains unknown.

View Article and Find Full Text PDF

Pulmonary vein anatomical variations are frequently observed in atrial fibrillation patients undergoing catheter ablation. However, when it comes to patients with atrial fibrillation and bilateral common ostium in the inferior pulmonary veins, using a bilateral circumferential pulmonary vein isolation approach during catheter ablation heightens the risk of esophageal injury. At present, there is no established standard catheter ablation strategy for such cases.

View Article and Find Full Text PDF

Objective This study aims to utilize the TriNetX database, a comprehensive global network, to improve our understanding of the frequency, demographic factors, and related comorbidities of surgical patients who develop venous thromboembolism (VTEs) events. Methods The global collaborative network in TriNetX was queried for all cases from January 1, 2017, through December 31, 2023. International Classification for Disease (ICD) diagnosis codes were used to define patient cohorts with deep vein thrombosis (DVT) of the upper or lower extremity or pulmonary embolism (PE).

View Article and Find Full Text PDF

There is limited available data regarding the impact of cycle length (CL) prolongation when converting atrial fibrillation (AF) to organized atrial tachycardia (AT) and its effect on clinical outcomes. We retrospectively screened and included a cohort of 132 patients with persistent or long-standing persistent AF who underwent circumferential pulmonary vein isolation (CPVI) and left atrial substrate modification (LASM) between January 2015 and October 2019. In all 132 consecutive patients, persistent AF was successfully converted into organized AT.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!