[Normalization methods for ethanol Raman spectra quantitative analysis].

Guang Pu Xue Yu Guang Pu Fen Xi

Institute of Biomedical Engineering, Jinan University, Guangzhou 510632, China.

Published: April 2010

By performing different Raman normalization methods for ethanol quantitative analysis, the authors proposed the method which used the highest band intensity of the ethanol Raman spectra in the maximum concentration as the normalization metric for ethanol concentration quantitative analysis. By using this method, the correlation coefficient was 0.999, the mean relative error was only 0.067 8, and the relative standard deviation (RSD) of the data from different experimental groups was 0.046 3. Both the validity and accuracy of this method were much better than the internal standard and ratio methods. Combined with baseline correction, the method can not only effectively resist the data fluctuation between different experimental groups, but also improve the accuracy of ethanol quantitative analysis obviously. The test using the method to measure the ethanol concentration of some wines from market proved that the relative standard deviations were all smaller than 0.012, indicating that the method is excellent for ethanol concentration quantitative analysis in commercial applications.

Download full-text PDF

Source

Publication Analysis

Top Keywords

quantitative analysis
16
ethanol concentration
12
methods ethanol
8
ethanol raman
8
raman spectra
8
ethanol quantitative
8
concentration quantitative
8
relative standard
8
experimental groups
8
ethanol
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!