The Virtual Intracranial Stenting Challenge 2007 (VISC'07) is becoming a standard test case in computational minimally invasive cerebrovascular intervention. Following views expressed in the literature and consistent with the recommendations of a report, the effects of non-Newtonian viscosity and pulsatile flow are reported. Three models of stented cerebral aneurysms, originating from VISC'07 are meshed and the flow characteristics simulated using commercial computational fluid dynamics (CFD) software. We conclude that non-Newtonian and pulsatile effects are important to include in order to discriminate more effectively between stent designs.