Purpose: The aim of the study was to determine the feasibility of using a clinical optical breast scanner with molecular imaging strategies based on modulating light transmission.

Procedures: Different concentrations of single-walled carbon nanotubes (SWNT; 0.8-20.0 nM) and black hole quencher-3 (BHQ-3; 2.0-32.0 µM) were studied in specifically designed phantoms (200-1,570 mm(3)) with a clinical optical breast scanner using four wavelengths. Each phantom was placed in the scanner tank filled with optical matching medium. Background scans were compared to absorption scans, and reproducibility was assessed.

Results: All SWNT phantoms were detected at four wavelengths, with best results at 684 nm. Higher concentrations (≥8.0 µM) were needed for BHQ-3 detection, with the largest contrast at 684 nm. The optical absorption signal was dependent on phantom size and concentration. Reproducibility was excellent (intraclass correlation 0.93-0.98).

Conclusion: Nanomolar concentrations of SWNT and micromolar concentrations of BHQ-3 in phantoms were reproducibly detected, showing the potential of light absorbers, with appropriate targeting ligands, as molecular imaging agents for clinical optical breast imaging.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4156007PMC
http://dx.doi.org/10.1007/s11307-010-0356-3DOI Listing

Publication Analysis

Top Keywords

clinical optical
16
optical breast
16
molecular imaging
12
imaging agents
8
agents clinical
8
breast imaging
8
breast scanner
8
optical
6
imaging
5
imaging light-absorbing
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!