Background: The characteristics of the ideal type of mesh are still being debated. Mesh shrinkage and fixation have been associated with complications. Avoiding shrinkage and fixation would improve hernia recurrence rates and complications. To our knowledge, this is the first study of a device with a self-expanding frame for laparoscopic hernia repair.
Methods: Six Rebound Hernia Repair Devices were placed laparoscopically in pigs. This device is a condensed polypropylene, super-thin, lightweight, macroporous mesh with a self-expanding Nitinol frame. The devices were assessed for adhesions, shrinkage, and histological examination. Laboratory and radiologic evaluations were also performed.
Results: The handling properties of the devices facilitated their laparoscopic placement. They were easily identified with simple x-rays. The mesh was firmly integrated within the surrounding tissue. One device was associated with 3 small adhesions. The other 5 HRDs had no adhesions. We noted no shrinkage or folding. All devices preserved their original size and shape.
Conclusions: At this evaluation stage, we found that the Rebound Hernia Repair Device may serve for laparoscopic hernia repair and has favorable handling properties. It prevents folding and shrinkage of the mesh. It may eliminate the need for fixation, thus preventing chronic pain. The Nitinol frame also allowed radiologic evaluation for gross movement. Further studies will be needed to evaluate its clinical application.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3021309 | PMC |
http://dx.doi.org/10.4293/108680810X12674612014824 | DOI Listing |
Surg Innov
January 2025
Division of General, Minimally Invasive, and Robotic Surgery, Department of Surgery, University of Illinois at Chicago, Chicago, IL, USA.
Background: Transabdominal pre-peritoneal inguinal hernia repair using the da Vinci Single-Port robot (SP-TAPP) is currently performed in few centers. We aimed to define the learning curve for SP-TAPP by analyzing operative times.
Methods: The operative times of 122 SP-TAPP performed between 2019 and 2024 were retrospectively analyzed.
Sci Rep
January 2025
Department of General Surgery, Shaoxing Central Hospital (The Central Affiliated Hospital, Shaoxing University), Shaoxing, 312030, Zhejiang Province, China.
Ventral hernias pose a prevalent challenge in abdominal wall surgery, with ongoing advancements in repair techniques designed to enhance patient outcomes. This study evaluates the efficacy, safety, and socio-economic impact of Totally Extraperitoneal Sublay Repair (TES) versus Laparoscopic Intraperitoneal Onlay Mesh Repair (IPOM) for small to medium-sized ventral hernias, with a particular focus on postoperative quality of life and patient satisfaction. A retrospective cohort study was conducted, encompassing 125 patients who underwent ventral hernia repair between May 2018 and November 2023.
View Article and Find Full Text PDFPolymers (Basel)
January 2025
Institute of Graduate Studies, Bioengineering Division, Tokat Gaziosmanpaşa University, 60250 Tokat, Türkiye.
Hernia repair is the most common surgical operation applied worldwide. Mesh prostheses are used to support weakened or damaged tissue to decrease the risk of hernia recurrence. However, the patches currently used in clinic applications have significant short-term and long-term risks.
View Article and Find Full Text PDFJ Clin Med
January 2025
Surgical Science Department, Plastic and Reconstructive Surgery Unit, "Pugliese-Ciaccio" Hospital, 88100 Catanzaro, Italy.
Incisional ventral hernia repair remains a challenging surgery for abdominal wall surgeons. We report the results at 48 months post-surgery regarding open ventral hernia repair (OVHR), analyzing the recurrence rate and incidence of chronic pain. This was a retrospective, observational study of 111 consecutive patients who underwent OVHR.
View Article and Find Full Text PDFSci Rep
January 2025
Department of Anesthesia and Pain Medicine, The First Affiliated Hospital of Chongqing Medical University, #1 Youyi Road, Yujiagang Community, Yuzhong District, Chongqing, 400016, China.
To optimize the efficacy of analgesia for patients receiving laparoscopic inguinal hernia repair, peri-operative pain neuroscience education (PNE) as a pain-specific cognitive therapy was incorporated into multi-modal analgesia. A randomized controlled trial was conducted to compare conventional analgesia (group CA) and the addition of pain neuroscience education into it (group PNE) in patients receiving laparoscopic inguinal hernia repair. Characteristics of peri-operative pain was evaluated with Douleur Neuropathique 4 questionnaire (DN-4), central sensitization inventory (CSI), pain catastrophizing scale (PCS) post-operatively and pressure pain threshold.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!