Introduction: Catheter ablation for paroxysmal atrial fibrillation is widely used for patients with drug-refractory paroxysms of arrhythmia. Recently, novel technologies have been introduced to the market that aim to simplify and shorten the procedure.
Aim: To compare the clinical outcome of pulmonary vein (PV) isolation using a multipolar circular ablation catheter (PVAC group), with point-by-point PV isolation using an irrigated-tip ablation catheter and the CARTO mapping system (CARTO group; CARTO, Biosense Webster, Diamond Bar, CA, USA).
Methods: Patients with documented PAF were randomized to undergo PV isolation using PVAC or CARTO. Atrial fibrillation (AF) recurrences were documented by serial 7-day Holter monitoring.
Results: One hundred and two patients (mean age 58 ± 11 years, 68 men) were included in the study. The patients had comparable baseline clinical characteristics, including left atrial dimensions and left ventricular ejection fraction, in both study arms (PVAC: n = 51 and CARTO: n = 51). Total procedural and fluoroscopic times were significantly shorter in the PVAC group (107 ± 31 minutes vs 208 ± 46 minutes, P < 0.0001 and 16 ± 5 minutes vs 28 ± 8 minutes, P < 0.0001, respectively). The AF recurrence was documented in 23% and 29% of patients in the PVAC and CARTO groups, respectively (P = 0.8), during the mean follow-up of 200 ± 13 days. No serious complications were noted in both study groups.
Conclusions: Clinical success rates of PV isolation are similar when using multipolar circular PV ablation catheter and point-by-point ablation with a three-dimensional (3D) navigation system in patients with PAF, and results in shorter procedural and fluoroscopic times with a comparable safety profile.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1540-8159.2010.02807.x | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!