Dependence of Yb-169 absorbed dose energy correction factors on self-attenuation in source material and photon buildup in water.

Med Phys

Radiation Laboratory, University of Massachusetts Lowell, 1 University Avenue, Lowell, Massachusetts 01854, USA.

Published: May 2010

Purpose: Absorbed dose energy correction factors, used to convert the absorbed dose deposited in a LiF thermoluminescent dosimeter (TLD) into the clinically relevant absorbed dose to water, were obtained for both spherical volumetric sources and for the model 4140 HDR Yb-169 source. These correction factors have a strong energy dependence below 200 keV; therefore, spectral changes were quantified as Yb-169 photons traveled through both source material (Yb2O3) and water with the corresponding absorbed dose energy correction factors, f(r, theta), calculated as a function of location in a phantom.

Methods: Using the MCNP5 Monte Carlo radiation transport simulation program, the Yb-169 spectrum emerging from spherical Yb2O3 sources (density 6.9 g/cm3) with radii between 0.2 and 0.9 mm were analyzed and their behavior compared against those for a point-source. The absorbed dose deposited to both LiF and H2O materials was analyzed at phantom depths of 0.1-10 cm for each source radius and the absorbed dose energy correction factor calculated as the ratio of the absorbed dose to water to that of LiF. Absorbed dose energy correction factors for the Model 4140 Yb-169 HDR brachytherapy source similarly were obtained and compared against those calculated for the Model M-19 Ir-192 HDR source.

Results: The Yb-169 average spectral energy, emerging from Yb2O3 spherical sources 0.2-0.9 mm in radius, was observed to harden from 7% to 29%; as these photons traveled through the water phantom, the photon average energy softened by as much as 28% at a depth of 10 cm. Spectral softening was dependent on the measurement depth in the phantom. Energy correction factors were found to vary both as a function of source radius and phantom depth by as much as 10% for spherical Yb2O3 sources. The Model 4140 Yb-169 energy correction factors depended on both phantom depth and reference angle and were found to vary by more than 10% between depths of 1 and 10 cm and angles of 0 degrees and 180 degrees. This was in contrast to that of the Model M-19 Ir-192 source which exhibited approximately 3.5%-4.4% variation in its energy correction factors from phantom depths of 0.5-10 cm. The absorbed dose energy correction factor for the Ir-192 source, on the other hand, was independent of angle to within 1%.

Conclusions: The application of a single energy correction factor for Yb-169 TLD based dosimetry would introduce a high degree of measurement uncertainty that may not be reasonable for the clinical characterization of a brachytherapy source; rather, an absorbed dose energy correction function will need to be developed for these sources. This correction function should be specific to each source model, type of TLD used, and to the experimental setup to obtain accurate and precise dosimetric measurements.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.3372291DOI Listing

Publication Analysis

Top Keywords

absorbed dose
44
energy correction
44
correction factors
32
dose energy
28
energy
14
correction
13
model 4140
12
correction factor
12
absorbed
11
dose
11

Similar Publications

The efficacy of photodynamic treatment (PDT) against deep-seated tumor is hindered by low penetration depth of light as well as hypoxic conditions which prevails in tumor. To overcome this limitation, Near-infrared (NIR) absorbing photosensitizers have been investigated actively. In the present study we evaluated the PDT efficacy of an NIR absorbing chlorophyll derivative 'Cycloimide Purpurin-18 (CIPp-18)' in Human Breast carcinoma (MCF-7) and cervical adenocarcinoma (Hela) cells under normoxic and hypoxic conditions.

View Article and Find Full Text PDF

Multi-layer shielding optimization of a high activity Am-Be mixed field irradiation facility.

Appl Radiat Isot

January 2025

Experimental Nuclear Physics Department, Nuclear Research Centre, Egyptian Atomic Energy Authority, Egypt; Cyclotron Facility, Egyptian Atomic Energy Authority, Egypt.

Neutron and gamma-ray shielding design for a 30Ci (1.11TBq) Am-Be irradiation facility is studied using MCNP5 Monte Carlo simulation code. The study focuses on the optimization of the shielding layers of the previously planned neutron irradiation facility.

View Article and Find Full Text PDF

Several exploratory studies have demonstrated the feasibility of cholecystokinin-2 receptor (CCK2R) targeting in patients with medullary thyroid carcinoma (MTC) and other neuroendocrine tumors (NETs). We report the results of a prospective phase I/IIA pilot study (clinicaltrials.gov NCT06155994) conducted at our center with the Ga-labeled peptide analog DOTA-DGlu-Ala-Tyr-Gly-Trp-(N-Me)Nle-Asp-1-Nal-Phe-NH (Ga-DOTA-MGS5).

View Article and Find Full Text PDF

Assessing fetal radiation dose from iodine-125 seeds in pregnant breast cancer patients: an updated model.

Phys Med Biol

January 2025

Department of Medical Physics, Jeroen Bosch Ziekenhuis, Henri Dunantstraat 1, 's-Hertogenbosch, 5223GZ, NETHERLANDS.

The treatment of breast cancer during pregnancy requires careful consideration of consequences for both maternal and fetal health. In non-pregnant patients, the use of radioactive iodine-125 (125I)-seeds is standard practice for localising non-palpable breast tumors before breast-conserving surgery. However, the use of 125I-seeds in pregnant patients has been avoided due to concerns about fetal radiation exposure.

View Article and Find Full Text PDF

Background And Objective: Psilocybin is currently being extensively studied as a potential therapeutic agent for multiple psychiatric disorders. Here, a systematic literature review of all published pharmacokinetic data on the pharmacologically active metabolite of psilocybin, psilocin, is presented.

Methods: The review includes clinical studies that reported pharmacokinetic data and/or parameters after psilocybin administration in humans.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!