Objective: To develop and test an arthroscopic aiming device for extra- to intraarticular tibial tunnel drilling emerging at the center of the tibial insertion (CenterTib) of the cranial cruciate ligament (CCL) in medium to large breed dogs.

Study Design: Descriptive experimental study.

Sample Population: Fifty-two cadaveric hind limbs of dogs >or=20 kg BW.

Methods: The mediolateral position and craniocaudal position (ccPos) of CenterTib in relation to the caudomedial meniscotibial ligament were measured on photographs of 46 dissected tibial plateaus. The proximal tibial depth (TibDepth) was determined on lateral radiographs and its correlation with ccPos was assessed using linear regression analysis. Extra- to intraarticular arthroscopic tibial tunnel drilling was performed in 6 independent cadaveric stifles. A C-guide with an adjustable craniocaudal offset was constructed and adjusted according to ccPos estimated based on TibDepth. The position of the resulting bone tunnels was compared with the position of the CenterTib.

Results: Pearson's correlation between TibDepth and ccPos was strong (R=0.86; P<.001). ccPos (y) as a function of TibDepth (x) can be expressed as y=-4.8+0.3x. Arthroscopic tunnel drilling resulted in a median deviation of the drill tunnels around the CenterTib of 1 mm.

Conclusion: The regression equation and aiming device permit localization and targeting of CenterTib during extra- to intraarticular tibial bone tunnel drilling in vitro.

Clinical Relevance: The proposed technique may reduce tibial tunnel misplacement when performing intraarticular CCL repair using a tibial bone tunnel.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1532-950X.2010.00652.xDOI Listing

Publication Analysis

Top Keywords

extra- intraarticular
12
tibial tunnel
12
tunnel drilling
12
aiming device
8
intraarticular tibial
8
cranial cruciate
8
cruciate ligament
8
tibial
6
development situ
4
situ application
4

Similar Publications

Functional alignment is a feasible alignment strategy in robotic assisted total knee arthroplasty for knee osteoarthritis with extra-articular deformity - A case series.

SICOT J

January 2025

Department of Orthopaedic Surgery, Joint Replacement Unit, Kuala Lumpur Hospital, Ministry of Health Malaysia, Jalan Pahang, 50586 Kuala Lumpur, Malaysia.

Introduction: Extraarticular deformity (EAD) with knee arthritis is a complex condition involving tri-planar bone deformity with pathological malalignment and chronic soft tissue contracture or laxity in the knee joint. Intraarticular correction by TKA, which was previously performed with conventional manual jig by mechanical alignment technique, had its limits and difficulties especially extensive soft tissue release and risk of jeopardizing the collateral ligaments. Robotic technology allows for reproducible and precise execution of surgical plan and allows adjustment to various new personalised alignment philosophy including functional alignment (FA).

View Article and Find Full Text PDF

Objectives: To report the frequency of patients with pre- and post-reduction computed tomography (CT) scans associated with acetabular fracture-dislocations and the change of associated intra-articular fragments occurring with joint reduction.

Methods: Design: Retrospective case series.

Setting: Regional Level 1 trauma center.

View Article and Find Full Text PDF

: The objective of our study was to evaluate the association and frequency of subcutaneous lymphedema in patients with gout primarily affecting the feet. : In 79 patients with acute gout, ultrasound (US) and dual-energy computed tomography (DECT) were performed to assess the presence of subcutaneous edema and extra- and intra-articular gouty deposits. In addition, the diagnostic utility of two post-processing DECT protocols were evaluated, comprising different minimum attenuation thresholds of 150 HU (DECT 150 protocol) and 120 HU (DECT 120 protocol), with the same maximum attenuation threshold (500 HU) and constant kilovoltage setting of tubes A and B at 80 and 140 kVp.

View Article and Find Full Text PDF

Background: Contamination of sterilized surgical instruments is not a typically suspected source of increased infection rate, especially if no abnormalities in the sterilization process are detected.

Purpose/hypothesis: The purpose of this study was to report increased infection rates after knee ligament reconstructions due to undetectable sterilization process errors leading to residual moisture, not limited to a specific surgical tool. It was hypothesized that (1) residual moisture on surgical tools due to autoclave overloading would not be detected by autoclave self-diagnostics, chemical and biological tests, or organoleptic assessment and (2) this kind of contamination may elevate infection rates, especially in knee intra-articular reconstruction procedures.

View Article and Find Full Text PDF

Introduction: The purpose of this study was to compare leg length discrepancies (LLD) between simultaneous and staged bilateral total hip arthroplasty (BTHA).

Materials And Methods: This retrospective cohort study included 150 consecutive patients who underwent BTHA (60 simultaneous and 90 staged BTHA). To adjust for preoperative patient characteristics between groups, 1:1 propensity score matching was performed based on age, sex, body mass index, American Society of Anesthesiologists Physical Status, Tönnis classification, Crowe classification, preoperative range of motion (flexion and abduction), LLD, and Harris Hip Score (HHS).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!