Purpose: The center method of corneal endothelial cell analysis is rapid but excludes the outermost digitized cells of a contiguous group from analysis; the flex-center method (Konan, Inc) is a modification that includes analysis of the outermost cells, which is advantageous in images with few cells. In this study, we examined agreement among the flex-center, center, and corner (standard) methods of endothelial analysis.
Methods: Identical cells in endothelial images of 10 normal corneas and 10 corneas after penetrating keratoplasty (PK) were analyzed by each method. Agreement among methods for endothelial cell density (ECD), coefficient of variation of cell area (CV), and the percentage of hexagonal cells (HEX) was assessed by using a Student-Newman-Keuls procedure.
Results: In normal corneas, there were small (clinically insignificant) differences among methods for ECD (P < 0.001) (mean +/- SD: flex center, 2846 +/- 248 cells/mm; center, 2870 +/- 253 cells/mm; and corner, 2892 +/- 254 cells/mm), CV (P < 0.001) (flex center, 33% +/- 3%; center, 30% +/- 3%; and corner, 30% +/- 3%), and HEX (P = 0.004) (flex center, 60% +/- 6%; center, 60% +/- 8%; and corner, 58% +/- 7%). In PK corneas, the methods agreed for ECD (P = 0.06) (flex center, 908 +/- 319 cells/mm; center, 912 +/- 307 cells/mm; and corner, 929 +/- 333 cells/mm) but disagreed for CV (P = 0.02) (flex center, 35% +/- 13%; center, 30% +/- 10%; and corner, 35% +/- 15%) and HEX (P = 0.02) (flex center, 56% +/- 19%; center, 54% +/- 17%; and corner, 43% +/- 23%).
Conclusion: ECD agreed among methods in normal and PK corneas, whereas morphometric data agreed poorly in PK corneas.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/ICO.0b013e3181cc7a60 | DOI Listing |
Comput Struct Biotechnol J
December 2024
Biological and Chemical Research Center, Faculty of Chemistry, University of Warsaw, Zwirki i Wigury 101, Warsaw 02-089, Poland.
CABS-flex is a well-established method for fast protein flexibility simulations, offering an effective balance between computational efficiency and accuracy in modeling protein dynamics. To further enhance its predictive capabilities, we propose incorporating AlphaFold's predicted Local Distance Difference Test (pLDDT) scores into CABS-flex simulations. The pLDDT scores, which reflect the confidence of AlphaFold's structural predictions, were integrated with secondary structure information to refine the restraint schemes used in the simulations.
View Article and Find Full Text PDFJ Appl Lab Med
December 2024
Department of Pathology, Johns Hopkins University School of Medicine, Baltimore, MD, United States.
Background: Pharmacogenomics has demonstrated benefits for clinical care, including a reduction in adverse events and cost savings. However, barriers in expanded implementation of pharmacogenomics testing include prolonged turnaround times and integration of results into the electronic health record with clinical decision support. A clinical workflow was developed and implemented to facilitate in-house result generation and incorporation into the electronic health record at a large academic medical center.
View Article and Find Full Text PDFCatheter Cardiovasc Interv
December 2024
Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, USA.
Background: Impella RP and RP Flex can provide right ventricular (RV) support in numerous clinical situations. Notably, the redesigned RP Flex allows for an internal jugular approach compared to the femoral venous approach with RP.
Aims: We investigated the type and frequency of device-related complications as reported in the US Food and Drug Administration (FDA) Manufacturer and User Facility Experience (MAUDE) for both Impella RP and RP Flex to see if there are any differences.
Prosthet Orthot Int
December 2024
University Medical Center Göttingen (UMG), Department of Trauma Surgery, Orthopedics and Plastic Surgery (CUOP), Göttingen, Germany.
Background: Energy storing and return (ESAR) prosthetic feet often provide high energy return at the cost of increased foot height, compromised appearance, and reduced flexibility. An ESAR foot, Ottobock Evanto, with a novel compact design has been introduced (novel foot).
Objective: To characterize user preference and benefits of a novel foot compared with the subject's previously worn (everyday foot) and comparator foot, Össur Pro-Flex XC.
ASAIO J
December 2024
From the Department of Thoracic and Cardiovascular Surgery, Kaufman Center for Heart Failure Treatment and Recovery, Heart, Vascular and Thoracic Institute, Cleveland Clinic, Cleveland, Ohio.
We present a case of a 56 year old man presenting with cardiogenic shock from isolated right ventricular failure due to arrhythmogenic right ventricular cardiomyopathy who was successfully bridged to heart transplantation after 29 days on an Impella RP Flex right ventricular assist device. This represents the longest reported duration of support with the device to date and highlights the utility of this platform in the management of severe right ventricular failure.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!