The laser integration line (LIL) located at CEA-CESTA is equipped with x-ray plasma diagnostics using different kinds of x-ray components such as filters, mirrors, crystals, detectors, and cameras. The CEA-DAM of Arpajon is currently developing x-ray calibration methods and carrying out absolute calibration of LIL x-ray photodetectors. To guarantee LIL measurements, detectors such as x-ray cameras must be regularly calibrated close to the facility. A new x-ray facility is currently available to perform these absolute x-ray calibrations. This paper presents the x-ray tube based high energy x-ray source delivering x-ray energies ranging from 0.9 to 10 keV by means of an anode barrel. The purpose of this source is mainly to calibrate LIL x-ray cameras but it can also be used to measure x-ray filter transmission of plasma diagnostics. Different x-ray absolute calibrations such as x-ray streak and framing camera yields, x-ray charge-coupled device quantum efficiencies, and x-ray filter transmissions are presented in this paper. A x-ray flat photocathode detector sensitivity calibration recently performed for a CEA Z-pinch facility is also presented.