Background: The CGMS Gold continuous glucose monitor presents a problem of lack of accuracy, especially in the lower range, sometimes leading to missed or false alarms. The new algorithm aims to improve the measurement accuracy and hypoglycemia detection.

Material/methods: Twenty-one patients with type 1 diabetes were monitored for 3 days (1 day at the hospital and 2 at home) using the CGMS Gold. For these patients, blood glucose samples were taken every 15 minutes for 2 hours after meals and every half hour otherwise during the first day. A new calibration algorithm was developed and implemented using CGMS Gold intensity readings and capillary glucose.

Results: After 1 day, a comparison of results from either the CGMS Gold algorithm and the proposed algorithm, compared with results from blood (2450 points), showed an increase of data in zone A with the proposed algorithm (4.4% in the Clarke error grid analysis (EGA) and 5.0% in the Consensus EGA). After comparing for 3 days, a reduction of 24.7%, p<0.05, in the overall median relative absolute difference (RAD) was also obtained. In the hypoglycemic range, a significant decrease in median RAD was observed (64.4%, p<0.05). Furthermore, the undetected hypoglycemia events in capillary samples by the proposed algorithm were reduced by 59.8% compared to the CGMS Gold algorithm.

Conclusions: The performance as measured with clinical and numerical accuracy criteria illustrates the improved accuracy of the proposed algorithm in comparison with the CGMS Gold algorithm. A significant improvement in hypoglycemia detection was observed.

Download full-text PDF

Source

Publication Analysis

Top Keywords

cgms gold
16
continuous glucose
8
proposed algorithm
8
algorithm
5
enhanced algorithm
4
glucose
4
algorithm glucose
4
glucose estimation
4
estimation continuous
4
glucose monitoring
4

Similar Publications

Objective: This study aimed to explore the relationship between time below range (TBR), impaired awareness of hypoglycemia (IAH), and severe hypoglycemia (SH).

Research Design And Methods: This cross-sectional study analyzed data from individuals with diabetes using continuous glucose monitors (CGMs) in the Association of British Clinical Diabetologists audit. Hypoglycemia awareness was assessed via the Gold score (≥4 denoting IAH), and SH was defined as hypoglycemia requiring third-party assistance.

View Article and Find Full Text PDF

Aims/hypothesis: We aimed to: (1) externally validate the five-item Hypoglycaemia Awareness Questionnaire (HypoA-Q) impaired awareness subscale (HypoA-Q IA); (2) examine how impaired awareness of hypoglycaemia (IAH) relates to the risk of severe hypoglycaemia and level 2 hypoglycaemia; and (3) identify factors associated with IAH.

Methods: Nationwide survey of T1D Exchange registrants was conducted to collect data on demographics, 6 month severe-hypoglycaemia history, hypoglycaemia awareness status (via HypoA-Q IA, the Gold instrument and the Clarke instrument) and continuous glucose monitor (CGM) measures. The Clarke hypoglycaemia awareness factor (Clarke-HAF) was calculated to exclude severe-hypoglycaemia history items.

View Article and Find Full Text PDF

Aims: To evaluate relationships of hypoglycemia awareness, hypoglycemia beliefs, and continuous glucose monitoring (CGM) glycemic profiles with anxiety and depression symptoms in adults with type 1 diabetes (T1D) who use CGM.

Methods: A cross-sectional survey and data collections were completed with 196 T1D adults who used CGM (59% also used automated insulin delivery devices (AIDs)). We assessed hypoglycemia awareness (Gold instrument), hypoglycemia beliefs (Attitudes to Awareness of Hypoglycemia instrument), CGM glycemic profiles, demographics, and anxiety and depression symptoms (Hospital Anxiety and Depression Scale).

View Article and Find Full Text PDF

This study aimed to evaluate the accuracy of Dexcom G6 (DG6) and FreeStyle Libre-2 (FSL2) during aerobic training and high-intensity interval training (HIIT) in individuals with type 1 diabetes. Twenty-six males (mean age 29.3 ± 6.

View Article and Find Full Text PDF

Aims: To evaluate relationships between hypoglycemia awareness, hypoglycemia beliefs, and continuous glucose monitoring (CGM) glycemic profiles and anxiety and depression symptoms in adults with type 1 diabetes (T1D) who use CGM or automated insulin delivery devices.

Methods: A cross-sectional survey and data collections were completed with 196 T1D adults who used advanced diabetes technologies. We assessed hypoglycemia awareness (Gold instrument), hypoglycemia beliefs (Attitudes to Awareness of Hypoglycemia instrument), CGM glycemic profiles, demographics, and anxiety and depression symptoms (Hospital Anxiety and Depression Scale).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!