Background: The CGMS Gold continuous glucose monitor presents a problem of lack of accuracy, especially in the lower range, sometimes leading to missed or false alarms. The new algorithm aims to improve the measurement accuracy and hypoglycemia detection.
Material/methods: Twenty-one patients with type 1 diabetes were monitored for 3 days (1 day at the hospital and 2 at home) using the CGMS Gold. For these patients, blood glucose samples were taken every 15 minutes for 2 hours after meals and every half hour otherwise during the first day. A new calibration algorithm was developed and implemented using CGMS Gold intensity readings and capillary glucose.
Results: After 1 day, a comparison of results from either the CGMS Gold algorithm and the proposed algorithm, compared with results from blood (2450 points), showed an increase of data in zone A with the proposed algorithm (4.4% in the Clarke error grid analysis (EGA) and 5.0% in the Consensus EGA). After comparing for 3 days, a reduction of 24.7%, p<0.05, in the overall median relative absolute difference (RAD) was also obtained. In the hypoglycemic range, a significant decrease in median RAD was observed (64.4%, p<0.05). Furthermore, the undetected hypoglycemia events in capillary samples by the proposed algorithm were reduced by 59.8% compared to the CGMS Gold algorithm.
Conclusions: The performance as measured with clinical and numerical accuracy criteria illustrates the improved accuracy of the proposed algorithm in comparison with the CGMS Gold algorithm. A significant improvement in hypoglycemia detection was observed.
Download full-text PDF |
Source |
---|
Diabetes Care
January 2025
Allam Diabetes Centre, Hull University Teaching Hospitals National Health Service (NHS) Trust, Hull, U.K.
Objective: This study aimed to explore the relationship between time below range (TBR), impaired awareness of hypoglycemia (IAH), and severe hypoglycemia (SH).
Research Design And Methods: This cross-sectional study analyzed data from individuals with diabetes using continuous glucose monitors (CGMs) in the Association of British Clinical Diabetologists audit. Hypoglycemia awareness was assessed via the Gold score (≥4 denoting IAH), and SH was defined as hypoglycemia requiring third-party assistance.
Diabetologia
February 2025
School of Psychology, Faculty of Health, Deakin University, Geelong, VIC, Australia.
Aims/hypothesis: We aimed to: (1) externally validate the five-item Hypoglycaemia Awareness Questionnaire (HypoA-Q) impaired awareness subscale (HypoA-Q IA); (2) examine how impaired awareness of hypoglycaemia (IAH) relates to the risk of severe hypoglycaemia and level 2 hypoglycaemia; and (3) identify factors associated with IAH.
Methods: Nationwide survey of T1D Exchange registrants was conducted to collect data on demographics, 6 month severe-hypoglycaemia history, hypoglycaemia awareness status (via HypoA-Q IA, the Gold instrument and the Clarke instrument) and continuous glucose monitor (CGM) measures. The Clarke hypoglycaemia awareness factor (Clarke-HAF) was calculated to exclude severe-hypoglycaemia history items.
Diabetes Res Clin Pract
March 2024
Department of Family Medicine, University of Michigan Medical School, Ann Arbor, MI 48105, USA.
Aims: To evaluate relationships of hypoglycemia awareness, hypoglycemia beliefs, and continuous glucose monitoring (CGM) glycemic profiles with anxiety and depression symptoms in adults with type 1 diabetes (T1D) who use CGM.
Methods: A cross-sectional survey and data collections were completed with 196 T1D adults who used CGM (59% also used automated insulin delivery devices (AIDs)). We assessed hypoglycemia awareness (Gold instrument), hypoglycemia beliefs (Attitudes to Awareness of Hypoglycemia instrument), CGM glycemic profiles, demographics, and anxiety and depression symptoms (Hospital Anxiety and Depression Scale).
Diabetes Technol Ther
June 2024
Clinimetry and Technological Development in Therapeutic Exercise Research Group (CLIDET), Department of Physiotherapy, University of Valencia, Valencia, Spain.
This study aimed to evaluate the accuracy of Dexcom G6 (DG6) and FreeStyle Libre-2 (FSL2) during aerobic training and high-intensity interval training (HIIT) in individuals with type 1 diabetes. Twenty-six males (mean age 29.3 ± 6.
View Article and Find Full Text PDFDiabetes Res Clin Pract
December 2023
Department of Family Medicine, University of Michigan Medical School, Ann Arbor, MI 48105, USA.
Aims: To evaluate relationships between hypoglycemia awareness, hypoglycemia beliefs, and continuous glucose monitoring (CGM) glycemic profiles and anxiety and depression symptoms in adults with type 1 diabetes (T1D) who use CGM or automated insulin delivery devices.
Methods: A cross-sectional survey and data collections were completed with 196 T1D adults who used advanced diabetes technologies. We assessed hypoglycemia awareness (Gold instrument), hypoglycemia beliefs (Attitudes to Awareness of Hypoglycemia instrument), CGM glycemic profiles, demographics, and anxiety and depression symptoms (Hospital Anxiety and Depression Scale).
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!